Chapter 7: Repair Procedures
Left Rail Internal Repairs
Service Manual
7-27
23.
Reinstall the connection panel at the bottom of the electrical enclosure. Fasten the control panel cover
to the back of the enclosure.
7.5 Left Rail Internal Repairs
7.5.1 Removing the Inner Rail Assembly
1.
Using a 2.5 mm hex key, remove the 2 screws in the rear dovetail of the upright that secure the cable
carrier channel to the inside of the upright. One screw is located at the top of the upright; the other is
half way down. Using a 3 mm hex key, remove the side roller bracket. Pull the entire rail assembly out of
the upright.
2.
Remove the pinch guard by sliding it up then pulling it off.
STEP 7
FIGURE 7-28. Left Inner Rail
Side roller bracket
Cable carrier
Cable carrier channel
Lift rail
Left upright
Summary of Contents for Giraffe OmniBed Carestation CS1
Page 4: ...RH 2 Service Manual ...
Page 24: ...Service Language Disclaimer viii Service Manual ...
Page 40: ...Safety and Regulatory Information User Responsibility xxiv Service Manual ...
Page 44: ...Chapter 1 Functional Description Control Board 1 2 Service Manual FIGURE 1 1 Block Diagram ...
Page 74: ...Chapter 3 Calibration and Maintenance Line Voltage Calibration 3 12 Service Manual ...
Page 82: ...Chapter 4 Understanding the Service Interface Upgrade 4 8 Service Manual ...
Page 110: ...Chapter 5 Troubleshooting Diagnostic Readings 5 28 Service Manual ...
Page 211: ...Chapter 7 Repair Procedures Display Harness Connections Service Manual 7 75 ...
Page 212: ...Chapter 7 Repair Procedures Display Harness Connections 7 76 Service Manual ...
Page 294: ...Chapter 8 Illustrated Parts Wiring Diagrams 8 82 Service Manual ...
Page 296: ...Appendix A Compartment and Skin Probe Characteristics A 2 Service Manual ...
Page 304: ...Appendix C RS 232 Serial Data Nurse Call C 4 Service Manual ...