11.5 Unit Cover / Control Board Replacement
SENSITIVE TO ELECTROSTATIC DISCHARGE CAUTION
This procedure includes ESD sensitive parts. ESD control guidelines must be followed during this
procedure to ensure that static charges are safely conducted to the ground and not through the
sensitive device, to prevent damage to the equipment.
1. Make sure the system power cord is unplugged from the power outlet.
2. Remove the light pipe as instructed in section 11.3.
3. Release the power cord retainer to unplug the power cord from the system.
4. Use a Philips screwdriver to remove the back plate screws (Refer to Figure 11-7).
5. Slide off the unit cover gently until the control board harness is accessible and disconnect the harness
from the LED driver board (Refer to Figure 11-8).
6. To remove the control board, use a Philips screwdriver to remove the control board screws and pull out
the fish paper, control board and the LCD cover.
7. Reverse steps to re-install.
NOTE:
When re-installing the unit cover, ensure the standby/on switch is aligned properly with its
corresponding hole on the unit cover prior to cover installation.
Figure 11-7
Figure 11-8
67
Chapter 11: Replacement Procedures
Summary of Contents for Giraffe Blue Spot PT Lite
Page 1: ...Giraffe Blue Spot PT Lite Operation Maintenance and Service Manual GE Healthcare ...
Page 7: ... 7 Part I Operation and Maintenance ...
Page 8: ...8 Part I Operation and Maintenance This page is intentionally left blank ...
Page 16: ...This page is intentionally left blank 16 Chapter 1 Product Overview ...
Page 22: ...This page is intentionally left blank 22 Chapter 2 Product Setup and Operation ...
Page 26: ...This page intentionally left blank 26 Chapter 3 Operator s Maintenance ...
Page 32: ...This page intentionally left blank 32 ...
Page 33: ... 33 Part II Service ...
Page 34: ...34 Part II Service This page is intentionally left blank ...
Page 44: ...44 Important Service Safety Information This page is intentionally left blank ...
Page 50: ...This page is intentionally left blank 50 Chapter 7 Installation ...
Page 55: ...Chapter 9 Calibration No calibration is required for the Giraffe Blue Spot PT Lite 55 ...
Page 56: ...This page is intentionally left blank 56 Chapter 9 Calibration ...
Page 72: ...This page is intentionally left blank 72 Chapter 11 Replacement Procedures ...
Page 88: ...12 5 Wiring Diagram 88 Chapter 12 Service Parts ...
Page 94: ...This page intentionally left blank 94 ...
Page 95: ......