Controls and Indicators
Page 6
2102322-111 REV 1
•
DAMAGE TO DEVICE AND ACCESSORIES
–
Unauthorized personnel do not have the proper training to
repair the device. Repairs carried out by unauthorized personnel could result in damage to the device
or accessories. Send the device for inspection to an authorized facility if you find or even suspect a
malfunction. Please add a detailed description of the observed malfunction.
•
DAMAGE TO THE DEVICE
–
Take care to prevent chemicals\liquids from entering the connectors or
internal part of the device.
•
PACEMAKER - It is recommended that a minimum separation of 15 cm (6 inches) be maintained
between the wireless patient unit 1200W and a pacemaker to avoid potential interference with
pacemaker. Some studies have shown that wireless devices might interfere with implanted cardiac
pacemakers if used within eight inches of the pacemaker. Pacemaker users may want to avoid placing
or using a wireless device this close to their pacemaker. If you have any reason to suspect that
interference is taking place, turn off the ECG immediately.
!
Caution
Use only while patient unit 1200W is worn on patient with its strap.
Use only with battery compartment of patient unit 1200W is closed.
NOTE
•
The GEH-ECG 1200 complies with Part 15 of the FCC Rules. Operation is subject to the following two
conditions:
This device may not cause harmful interference and
This device must accept any interference received, including interference that may cause
undesired operation.
•
The manufacturer is not responsible for any Radio or TV interference caused by unauthorized
modifications to this equipment. Such modifications could void the user’s authority to operate the
equipment.
NOTE
NO OPTIMUM WIRELESS PERFOMANCE
Note that the complete data transfer cannot be guaranteed under all circumstances when using Norav
proprietary RF technology and that range loss can occur if this technology and other RF Devices (Bluetooth,
WLAN.) are used near each other. Also, other devices could interfere with the equipment, even if the other
devices comply with CISPR emission requirements
In case of interruptions with the Norav proprietary RF technology:
•
Remove any objects from the line of sight between patient acquisition unit 1200W and Receiver
unit 1200WR
•
Remove any possible RF interfering devices from the vicinity of 1200W and 1200WR.
•
Do not position the devices closer than 50cm to each other or more than 10m from each other.
Controls and Indicators
Element
Description
1
Patient unit 1200W – Power ON/OFF Button
2
Patient unit 1200W – Low Battery indicator
3
Patient unit 1200W – Power ON/OFF indicator. Transmit/Standby indicator
4
Patient unit 1200W – Detachable ECG Leadwires
5
Patient unit 1200W – Leadwires guides. Strap Holders.
6
Patient unit 1200W –
Battery compartment door
7
Receiver unit 1200WR – Detachable RF Antenna
8
Receiver unit 1200WR – Power ON/OFF indicator.
9
Receiver unit 1200WR – Detachable USB Cable
10
Receiver unit 1200WR – Auxiliary BNC output