Performing a Quality Control Test
1. Take one GE006 Blood Glucose Test Strip from the vial and close
the vial cap immediately.
2. Insert the GE006 Blood Glucose Test Strip horizontally (not
vertically) with the colored view window facing up into the test
strip port.
3. While the blood drop icon is flashing on the display window, press
and hold the Main button for at least 3 seconds until the " "
symbol appears.
4. You will see the blinking "
" icon and " " icon on the
screen, prompting you to apply
.
5. Shake the bottle of
well before
opening the cap. Place the cap on a flat surface.
6. Place a drop of control solution onto the top of the cap.
7. Gently touch the sample port of the strip with the control solution
from the top of the cap.
8. The screen will display the count time starting from 5 (you will
hear a beep if the volume is turned on).
Rightest™
Rightest™
Control Solution
GC550
Control Solution GC550
8
OPEN
5
6
1
2
3
Performing a Quality Control Test
9. Tightly replace the cap on the
bottle.
10. The control solution result will appear. Compare your Quality Control Test result
to the Control Solution Range printed on the GE006 Blood Glucose Test Strip vial
label.
Rightest™
Control Solution GC550
CLOSE
9
CAUTION
- Your Control Solution Test results will not be included in the average calculations, however,
they can still can be recalled and viewed. The Control Solution Test result will be shown with
the " " icon on the screen.
- The Control Solution Test should be conducted between 15-40°C (59-104°F).
- Before " " and " " appear, do not touch the control solution to the sample port on the
strip. The GE006 Blood Glucose Meter is performing an internal check. Touching the control
solution to the sample port before prompted will result in an error message: " " and
“ "
.
- Do not drip the control solution on to the sample port of the test strip directly.
The reagent on the strip might leak into the bottle of control solution and may
cause the degeneration of the control solution. This could contaminate the meter
via the test strip port.
- Keep the test strip port clean and dry. Clean immediately if the test strip port is
stained or is overly exposed to moisture.
- Do not touch the tip of the control solution bottle. If the tip is touched, clean with water.
and be accompanied by beeps (if volume is turned on)