8-6-2-5
Install the 17” LCD Monitor
Follow these steps to install the LCD Monitor:
1.) Place the LCD assembly onto the LCD Arm Bracket. Be sure that the notches in the LCD fixing
bracket is well aligned onto the corresponding positions on the LCD Arm Bracket.
2.) Install the two screws (M5x8, Torque: 5.7 Nm).
3.) Tilt the monitor to horizontal position.
Figure 8-81 Install the LCD Monitor
Figure 8-82 Two fixing screws
NOTCHES
(ONE ON EACH SIDE
OF THE BRACKET,
ONE SHOWN HERE)
FIXING SCREWS