GE
D
IRECTION
FQ091019, R
EVISION
2
V
IVID Q
N S
ERVICE
M
ANUAL
8-18
Section 8-2 - External Component Replacement Procedures
3.) Lift and tilt the ECG Module on its side, then disconnect the ECG Module cable - see
.
4.) Carefully remove the ECG module assembly (board and bracket) from the base of the Control Panel
and place it on a table.
5.) Remove the tape from the Flex Cable on the ECG Board and disconnect the Flex Cable.
6.) Unfasten the two screws that secure the bracket to the ECG Board and remove the bracket.
7.) Fit the bracket to the new ECG/Respiratory Board, ready for the installation procedure.
8-2-4-5-10
ECG/Respiratory Module Assembly Installation Procedure
1) Carefully tilt the Control Panel and Keyboard on its side.
2.) Position a new ECG/Respiratory Module assembly in the correct location on the base of the Control
Panel.
3.) Re-connect the ECG cable - see
4.) Re-connect the Flex Cable (previously removed from the ECG Board) to the new ECG/Respiratory
Board.
5.) Return and fasten the four screws (two long and two short) on either side of the bracket to secure
the ECG/Respiratory Module assembly to the base of the Control Panel (refer to
). When performing this step, each long screw is inserted through one of the two
standoffs.
6.) Install the Control Panel and Keyboard, as described in the
Installation Procedure" on page 8-12
7.) Turn ON power to the system.
The system re-boots into Normal Scanning mode.
NOTE:
In the event that an error message is displayed prompting for deletion of the traces file, this
indicates failure to perform the Preparation procedure prior to installation of the new
ECG/Respiratory board. Refer to
.
Figure 8-22 Disconnecting the ECG Module Cable
Release
cable
here