NOTE:
Beneath one of the four screws removed in the previous step, is a small plastic securing tab
(illustrated in
) that serves to seat the BEP firmly in position. It is necessary to loosen
and release this tab (lift it up using a screw driver if necessary) in order to facilitate complete
removal of the BEP.
9) Using both hands, lift the BEP out of the Bottom Assembly and remove it from the system, as shown
Figure 8-47 Removing the Screws from the BEP
Figure 8-48 Releasing the Plastic Securing Tab from the BEP