
8
- [EN] English
K405 Issue 3
Operation - Measure/supply Volts or mV
To measure/supply Volts or mV:
1. Connect the instrument (Figure 4/5) and, if necessary,
adjust the
Set Up
2. Select the task from the task selection menu
Note: Use the Dual Function area (
➁
) to do two operations
at the same time. If the area
➁
selection is not necessary,
set this area to off
(
■
)
. This saves the battery power.
3. If necessary, adjust the
Settings
(Table 4/5) and/or the
output values to the system (Table 7).
Figure 4:
Example configuration - To measure/supply Volts
or mV (Area
➀
)
Figure 5:
Example configuration - To measure Volts (Dual
Function, area
➁
)
Operation - Measure/supply Hz or pulses
To measure/supply Hz or pulses:
1. Connect the instrument (Figure 6) and, if necessary,
adjust the
Set Up
2. Select the task from the task selection menu (Table 2):
3. If necessary, adjust the
Settings
(Table 4/5) and/or the
output values to the system (Table 7).
Figure 6:
Example configuration - To measure/supply Hz
or Pulses
For an input, the display shows the condition of the
frequency gate:
Operation - RTD/Ohms connections
In the examples that follow 2W, 3W, and 4W identify the 2,
3, and 4-wire connections for a RTD or resistance.
0 to ±30V OR
0 to ±120mV
Edit
Settings
a) Measure
b) Supply
0 to 12V OR
0 to 120mV
0 to ±30V
=
Gate open (measurement starts)
=
Gate closed (measurement is waiting for the next
rising edge of the cycle)
=
Fast cycle
Settings
Edit
Settings
a) Input
b) Output