Procedures (CH1): True Ohms (measure)
8.9
Do the procedure as follows:
1.
Repeat procedure 8.8; in step 3 and 4 select True Ohms.
2.
Check the error is within limits (Ref: Table 8-1).
Table 8-12 True Ohms (measure) error limits
Standard
Resistor
(
Ω
)
Resistor
uncertainty
(
Ω
)
Permitted
DPI 620 error
(
Ω
)
0 (short
circuit)
-
0.004
100
0.002
0.0095
200
0.004
0.015
300
0.006
0.0205
400
0.008
0.026
1000
0.02
0.059
2000
0.04
0.114
4000
0.08
0.224