Table 8-4 Millivolts (measure) error limits
Applied
mV
Calibrator
uncertainty
(mV)
Permitted
DPI 620 error
(mV)
±2000
0.051
0.14
±1000
0.040
0.1
±200
0.0051
0.017
±100
0.0040
0.0125
0 (short
circuit)
0.0036
0.008
Table 8-5 Voltage (measure) error limits
Applied
V
Calibrator
uncertainty
(V)
Permitted
DPI 620 error
(V)
±30
0.00052
0.0021
±21
0.0004
0.0018
±20
0.00031
0.0009
±10
0.00016
0.00065
±5
0.00008
0.00053
0
0.000024
0.0004
06 June 2013
K0541 issue 1
Page 147 of 175