Chapter 4. Operation
46 | ADTS 405 R/F Mk2 Instruction Manual-English
© 2017 General Electric Company – All Rights Reserved.
a.
Press and hold down
F1
then press
SETUP
to enter the configuration menu. Release
both keys.
b.
If requested, enter PIN details.
c.
Press [MORE], [DISPLAY/OPTIONS], [OPTIONS], [IEEE-488], [SCPI/OPT2].
d.
Check the display shows that SCPI is selected. If the settings are correct, press
CLEAR/QUIT
until the display shows the main pressure display.
Note:
When enabled, the IEEE 488 facility will be available from power-up of the
ADTS 405 Mk2. The setting is held in non-volatile memory.
4.14.2 IEEE 488 Connector
The IEEE 488 connector is located behind the (optional) communications module cover plate.
The connector’s pin-out is detailed in Table 4-4:
INFORMATION
The cover plate for communications option must
always be fitted when the equipment is used outdoors, or in an
environment where moisture may enter the unit through the connector
panel.
Table 4-4: IEEE 488 Connector Pin-out
Pin Number
Function
1
D101
2
D102
3
D103
4
D104
5
E01
6
DAV
7
NRFD
8
NDAC
9
IFC
10
SRQ
11
ATN
12
Shield
a
13
D105
14
D016
15
D107
16
D108
17
REN
18
GND
b
19
GND
b