4-4
DINAMAP
®
ProCare Monitor Service Manual
Revision C
2009381-001
Maintenance:
Cleaning
Cleaning
Cleaning the Monitor
CAUTIONS
Never pour or spray water or any cleaning solution on the
equipment or permit fluids to run behind switches, into
connectors, into the recorder, into the temperature probe
well, or into any ventilation openings in the equipment.
Do not let fluid “pool” around connection pins.
Never immerse monitor or accessories in any liquid.
Do not attach the monitor or accessories to a patient until
it is thoroughly dry.
WARNING
Failure to follow these cleaning recommendations may melt,
distort, or dull the finish of displays and cases; blur lettering on
labels; embrittle cases and lead to cracks and breakage; or
cause equipment failures. Use of non-approved cleaning
agents is not considered normal wear and repair or
replacement of parts is not covered under warranty.
Cleaning Recommendations
Monitor Exterior
Disconnect the monitor from AC power before cleaning or disinfecting its
surface. The exterior surfaces of DINAMAP Monitors may be cleaned with a
dampened, lint-free cloth. Wipe off all cleaning solutions with a clean, dry cloth
and let air dry for at least 15 minutes. Use one of the following approved
solutions:
Mild soap (diluted)
Commercial diluted bleach solution or bleach wipe
Commercial diluted ammonia solution
10% solution of household bleach (5.25% sodium hypochlorite) in distilled
water
Never use the following cleaning agents:
Abrasive cleaners or solvents of any kind
Acetone
Ketone
Betadine
Alcohol- or petroleum-based cleaning agents
Summary of Contents for DINAMAP ProCare
Page 2: ......
Page 11: ...Revision C DINAMAP ProCare Monitor Service Manual 1 1 2009381 001 1 Introduction ...
Page 12: ...1 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 For your notes ...
Page 23: ...Revision C DINAMAP ProCare Monitor Service Manual 2 1 2009381 001 2 Equipment Overview ...
Page 24: ...2 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 For your notes ...
Page 39: ...Revision C DINAMAP ProCare Patient Monitor Service Manual 3 1 2009381 001 3 Installation ...
Page 40: ...3 2 DINAMAP ProCare Patient Monitor Service Manual Revision C 2009381 001 For your notes ...
Page 53: ...Revision C DINAMAP ProCare Monitor Service Manual 4 1 2009381 001 4 Maintenance ...
Page 54: ...4 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 For your notes ...
Page 71: ...Revision C DINAMAP ProCare Monitor Service Manual 5 1 2009381 001 5 Troubleshooting ...
Page 72: ...5 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 For your notes ...
Page 78: ...6 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 For your notes ...
Page 99: ...Revsion C DINAMAP ProCare Monitor Service Manual A 1 2009381 001 A Technical Specifications ...
Page 100: ...A 2 DINAMAP ProCare Monitor Service Manual Revision C 2009381 001 For your notes ...
Page 133: ......