Foundation Fieldbus Communications
D-5
March 2008
Resetting Instrument
Totalizers
To reset the instrument totalizers:
1.
Double click on the
FLOW
transducer block (in the tree under
GEFlow-XMT
; see Figure D-1 on page D-2).
2.
Select the
Others
tab and scroll down to the
CLEAR_TOTALIZERS
listing.
3.
Select
Clear
from the drop down list box (see Figure D-4 below).
4.
After the totals have been reset, select
Normal
from the drop
down list box to resume total accumulation.
Figure D-4: Clear Totalizers Drop Down List