Calibration
1-5
April 2008
Calibrating the Analog
Outputs (cont.)
If the linearity test readings are not within 5
μ
A of the values listed in
Table 1-1 above, check the accuracy and wiring of the ammeter.
Then, repeat the low and high end calibrations. If the analog output
still does not pass the linearity test, contact the factory for assistance.
Procedure Options
You have completed calibration of the analog output. Do one of the
following:
•
To calibrate or test another output, press
[F4] (EXIT)
and return to
Step 3 in
Accessing the Calibration Menu
•
To calibrate/test additional slots, press
[F4] (EXIT)
twice and
proceed to the appropriate section.
•
To leave the Calibration Menu, press
[F4] (EXIT)
three times.
Table 1-1: Expected Ammeter Readings
% Full Scale
4-20 mA Scale*
0-20 mA Scale*
0
4.000
0.000
10
5.600
2.000
20
7.200
4.000
30
8.800
6.000
40
10.400
8.000
50
12.000
10.000
60
13.600
12.000
70
15.200
14.000
80
16.800
16.000
90
18.400
18.000
100
20.000
20.000
* All ammeter readings should be ±0.005 mA