March 2008
2-6
Error Codes and Screen Messages
Screen Messages
A variety of messages may appear on the display screen during the
performance of a task. Since the error codes have already been
discussed in this chapter and the locator bar messages are discussed in
detail in Chapter 3,
Operation
, of the
Startup Guide
, they will not be
repeated here. All other messages are listed in Table 2-1 below.
Table 2-1: Screen Messages
Message
Meaning
Power Up Messages
Backup Battery FAIL
The backup battery that powers the non-volatile RAM has failed. Contact
the factory.
Backup Battery OK
The backup battery that powers the non-volatile RAM has passed.
DSP Processor OK
The DSP (digital signal processor) has passed.
DSP RAM Failed
The DSP (digital signal processor) RAM has failed. Contact the factory.
DSP RAM OK
The DSP (digital signal processor) RAM has passed.
EPROM sum = XXXX
Record the EPROM sum at initial power up and periodically thereafter.
FRIGID_INIT Executed
The NVR (non-volatile RAM) was automatically initialized due to a memory
fault. Contact the factory.
NVR FAIL
The non-volatile RAM has failed. Contact the factory.
NVR OK
The non-volatile RAM has passed.
Measurement Mode Messages
All Logs Cleared! hit key
This message may appear during one of the following tasks:
1. clearing a log - there are no more logs to clear
2. logging data - user has hit [ENT] instead of selecting an old log or
entering a new name
3. printing a log - there are no logs to print
4. displaying a log - there are no logs to display
Hit any key to resume taking measurements.
All Sites Cleared! hit key
This message may appear during one of the following tasks:
1. saving a new site - a new site name was not entered or overwrite
existing site was not selected
2. recalling a site - there are no site files to recall
3. clearing a site - there are no sites to delete
4. printing a site - there are no sites to print
5. displaying a site - there are no sites to display
Hit any key to resume taking measurements.
Do you want to SAVE?
This message appears upon exiting the
User Program
, if the new site data
has not been saved. Failure to save will result in loss of the new data next
time the site data is changed or recalled.
Duplicate name,
Enter another.
The site file or log name is already in use. Enter a different name.
End Time must exceed
Start Time by 5 min.
This message appears when in the LOG menu. Enter an end time that is at
least five minutes later than the start time.
Header invalid
An option card error indicating a programming failure or a loss of memory.
Contact the factory.