Chapter 3. Diagnostics
Chapter 3. Diagnostics
3.1 Introduction
This chapter explains how to troubleshoot the Model DF868 if problems arise with the electronics console, the
flowcell, or the transducers. Indications of a possible problem include:
•
display of an error message on the active display screen
•
erratic flow readings
•
readings of doubtful accuracy (i.e., readings that are not consistent with readings from another flow measuring
device connected to the same process).
If any of the above conditions occurs, proceed with the instructions presented in this chapter.
3.2 Displaying Diagnostic Parameters
The Model DF868 has a built-in
Diagnostics Menu
to aid in the troubleshooting of flowcell, transducer and/or
electrical problems. The
Diagnostics Menu
may only be entered from the
Big
or
Dual
measurement mode display. See
Chapter 2,
Displaying Data,
in the
Programming Manual
for instructions on setting the display screen to the desired
format, and enter the
Diagnostics Menu
as shown below.
Note:
For a 1-Channel Model DF868, the above prompt does not appear, and the following prompt is the initial
screen.
Note:
Ch1 (or Ch2), which is shown in parentheses above, appears in the locator bar only with a 2-Channel Model
DF868.
(Ch1) label SITE
Press the [
]
or [
]
keys until the desired channel option appears on the option
bar. Press the appropriate
[Fx]
key to select this option.
Velocity Ft/s
6.95
(error codes appear here)
CH1
CH2
SUM
DIF
(Ch1) label SITE
Press the [
]
or [
]
keys until
DIAG
appears on the option bar. Press the
appropriate
[Fx]
key to select this option.
Velocity Ft/s
6.95
(error codes appear here)
TIME
DIAG