12 Diamond X-ray tube data sheets
86500-IMG rev 3
GE Healthcare
89
12.6 G
ENERAL
DATA
12.7 B-112 H
OUSING
Maximum Peak Voltage:
39 kV (Anode to Ground)
Nominal Focal Spot Size:
Small 0.1 mm (Refer to General
Information Section for Focal Spot
Measurement Method)
Large 0.3 mm
Maximum Anode Load:
3-phase, 180Hz, 0.1 second
Small Focal Spot:
1320 W
Large Focal Spot:
2.6 kW
Anode Disc Construction:
Molybdenum Alloy
Anode Disc Diameter:
77 mm
Anode Disc Target Angle:
M113SP (LF/SF) 16/10
Anode Heat Storage:
300 kHU
Maximum Anode Cooling Rate:
60,000 HU/minute
Inherent Filtration:
0.0 mm Alum. Equiv. 0.76 mm
Beryllium
U/L Approved:
Yes
Class:
I
Type:
B
Maximum Peak Voltage:
52 kV
Anode to Ground:
52 kV
Cathode to Ground:
0 kV
Housing Heat Storage Capacity:
370 KJ (500 kHU)
Housing Cooling Rate:
100 Watts (135HU/sec.)
Weight:
10.2 kg (22.5 lbs)
Length:
33 cm (13”)
Diameter:
13.8 cm (5.44”)
Focal Spot - Port Spacing:
Small 47.0mm ± 1mm
Large 41.4 ± 1mm
High Voltage Cable, Special Design
(Non-Federal Standard):
55 kV Rating
Acoustic noise:
50dB Max. (A-weighted) at 30 cm
Leakage Technique Factor M113SP: 52 kV, 2mA
Leakage Technique Factor B112:
55 kV, 4mA
Minimum Ambient temperature for
storage & transportation:
-9°C (16°F)
Maximum Operation Temperature: +40°C (+104°F)