Revision A
Dash 3000/4000 Patient Monitor
5-15
2000966-339
Troubleshooting:
Power Source Tests
2. Use a DMM to measure the ground-to-neutral loop resistance.
CAUTION
DO NOT check the ground-neutral loop resistance unless
the wall receptacle is correctly wired.
a. Select the milliohms (m
Ω
) scale on the DMM.
b. Measure resistance across the power cord ground and neutral.
c. Measure from the ground lug on the rear power connector to any
exposed metal of the monitor. The resistance between the ground
and neutral connections, after the ohmmeter is nulled, must be
less than 100 m
Ω
. If not, have the wall receptacle checked by an
electrician.
Power Cord and Plug
Verify the power cord being used with the monitor is good. The following
are a couple of things to check for in this regard:
Failure of the power cord strain relief is very common. Often times
users of the equipment pull on the power cord itself, rather than the
power cord plug, to unplug the monitor from a wall receptacle. If in
doubt, test for continuity through each conductor of the power cord
connector and plug.
Verify line, neutral, and ground conductors are properly connected to
the power cord plug and are not short-circuited. Rewire and tighten
these, or replace the power cord, as necessary.
Summary of Contents for Dash 3000
Page 1: ...Dash 3000 4000 Patient Monitor Service Manual 2023896 023 Revision A ...
Page 11: ...Revision A Dash 3000 4000 Patient Monitor 1 1 2000966 339 1 Introduction ...
Page 12: ...1 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 19: ...Revision A Dash 3000 4000 Patient Monitor 2 1 2000966 339 2 Equipment Overview ...
Page 20: ...2 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 45: ...Revision A Dash 3000 4000 Patient Monitor 3 1 2000966 339 3 Installation ...
Page 46: ...3 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 55: ...Revision A Dash 3000 4000 Patient Monitor 4 1 2000966 339 4 Maintenance ...
Page 56: ...4 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 117: ...Revision A Dash 3000 4000 Patient Monitor 5 1 2000966 339 5 Troubleshooting ...
Page 118: ...5 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 149: ...Revision A Dash 3000 4000 Patient Monitor 6 1 2000966 339 6 Configuration ...
Page 150: ...6 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 173: ...Revision A Dash 3000 4000 Patient Monitor 7 1 2000966 339 7 Calibration ...
Page 174: ...7 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 182: ...8 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 226: ...9 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 280: ...A 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Page 291: ......