4-4
Dash™ 3000/4000/5000
2000966-386D
Alarms
You can change patient status alarm levels and limits. For more information, refer to
.
System Status Alarms
System status alarms are triggered by mechanical or electrical problems. They are of
lesser priority than patient status alarms.
There are three levels of system status alarms:
WARNING
— Serious mechanical or electrical problems.
ADVISORY
— Mechanical or electrical problems.
MESSAGE
— Additional information only.
The monitor’s response to system status alarms is as follows.
NOTE
System status alarms cannot, in most cases, be moved from one level to another.
However, you can set default alarm levels for the following system status alarms:
ECG LEADS FAIL
,
SPO2 PROBE OFF,
SPO2 PULSE SEARCH
,
NBP NO
DETERMINATION
.
The System status alarm levels are configurable in
MONITOR DEFAULTS
>
SETUP DEFAULT DISPLAY
.
Indicator
CRISIS
WARNING
ADVISORY
MESSAGE
Alarm tone
Three beeps
Two beeps
One beep
No
Alarm light
Red
Yellow
No
No
On-screen message
Yes
Yes
Yes
Yes
Automatic print
1
1
The factory default for ADULT-ICU is ON; NEONATAL-ICU and OPERATING ROOM is OFF.
Yes
Yes
No
No
ALARM HISTORY
Yes
Yes
Yes
No
Alarm broadcast
Yes
Yes
Yes
No
Remote alarm terminal
Yes
Yes
Yes
No
Indicator
WARNING
ADVISORY
MESSAGE
Alarm tone
Repeating foghorn
Single foghorn
No
Alarm light
Yellow
No
No
On-screen message
Yes
Yes
Yes
Alarm broadcast
Yes
No
No
Remote alarm
terminal
Yes
No
No
Summary of Contents for Dash 3000
Page 11: ...2000966 386D Dash 3000 4000 5000 ix Symbols F 8 ...
Page 12: ...x Dash 3000 4000 5000 2000966 386D ...
Page 15: ...2000966 386D Dash 3000 4000 5000 1 1 1 Introduction ...
Page 33: ...2000966 386D Dash 3000 4000 5000 2 1 2 Equipment Overview ...
Page 48: ...2 16 Dash 3000 4000 5000 2000966 386D Equipment Overview ...
Page 49: ...2000966 386D Dash 3000 4000 5000 3 1 3 Monitor Setup ...
Page 61: ...2000966 386D Dash 3000 4000 5000 4 1 4 Alarms ...
Page 69: ...2000966 386D Dash 3000 4000 5000 5 1 5 Managing Patients ...
Page 81: ...2000966 386D Dash 3000 4000 5000 6 1 6 Patient Data ...
Page 92: ...6 12 Dash 3000 4000 5000 2000966 386D Patient Data ...
Page 93: ...2000966 386D Dash 3000 4000 5000 7 1 7 Printing ...
Page 97: ...2000966 386D Dash 3000 4000 5000 8 1 8 Monitoring ECG ...
Page 116: ...8 20 Dash 3000 4000 5000 2000966 386D Monitoring ECG ...
Page 117: ...2000966 386D Dash 3000 4000 5000 9 1 9 Monitoring Invasive Pressures ...
Page 125: ...2000966 386D Dash 3000 4000 5000 10 1 10 Monitoring NBP ...
Page 135: ...2000966 386D Dash 3000 4000 5000 11 1 11 Monitoring SpO2 ...
Page 149: ...2000966 386D Dash 3000 4000 5000 12 1 12 Measuring Cardiac Output ...
Page 155: ...2000966 386D Dash 3000 4000 5000 13 1 13 Monitoring Respiration ...
Page 159: ...2000966 386D Dash 3000 4000 5000 14 1 14 Monitoring Temperature ...
Page 162: ...14 4 Dash 3000 4000 5000 2000966 386D Monitoring Temperature ...
Page 163: ...2000966 386D Dash 3000 4000 5000 15 1 15 Monitoring End Tidal CO2 ...
Page 170: ...15 8 Dash 3000 4000 5000 2000966 386D Monitoring End Tidal CO2 ...
Page 171: ...2000966 386D Dash 3000 4000 5000 16 1 16 Monitoring Anesthesia Gases ...
Page 179: ...2000966 386D Dash 3000 4000 5000 17 1 17 Monitoring Impedance Cardiography ICG ...
Page 185: ...2000966 386D Dash 3000 4000 5000 18 1 18 Monitoring Bispectral Index BIS ...
Page 193: ...2000966 386D Dash 3000 4000 5000 A 1 A Maintenance ...
Page 210: ...A 18 Dash 3000 4000 5000 2000966 386D Maintenance ...
Page 211: ...2000966 386D Dash 3000 4000 5000 B 1 B Software Packages ...
Page 214: ...B 4 Dash 3000 4000 5000 2000966 386D Software Packages ...
Page 215: ...2000966 386D Dash 3000 4000 5000 C 1 C Factory Defaults ...
Page 227: ...2000966 386D Dash 3000 4000 5000 D 1 D Custom Defaults ...
Page 236: ...D 10 Dash 3000 4000 5000 2000966 386D Custom Defaults ...
Page 237: ...2000966 386D Dash 3000 4000 5000 E 1 E Interfacing with Non GE Devices ...
Page 255: ...2000966 386D Dash 3000 4000 5000 F 1 F Abbreviations and Symbols ...
Page 268: ...Index 6 Dash 3000 4000 5000 2000966 386D ...
Page 269: ......