![GE Dash 3000 Operator'S Manual Download Page 65](http://html.mh-extra.com/html/ge/dash-3000/dash-3000_operators-manual_128578065.webp)
Alarms
2000966-386D
Dash™ 3000/4000/5000
4-5
NOTE
Patient
CRISIS
,
WARNING
, and
ADVISORY
level alarms are higher priority
than System
WARNING
and System
ADVISORY
level alarms. For example,
with
LEADS FAIL
set to System
WARNING
, if a
LEADS FAIL
alarm occurs
while a Patient
ADVISORY
level alarm is active, the audible alarm associated
with the Patient
ADVISORY
level alarm will sound at the bedside monitor. In
addition, the System
WARNING
alarm will not activate a remote alarm terminal
or View On Alarm function. Visual alarms are still present on the bedside
monitor and any central station monitoring device.
Safety
WARNING
ALARM ACTIVATION—No alarms sound or display on the
monitor until a patient is admitted to the monitor. The monitor will
NOT alarm if an unadmitted patient enters an alarm condition. You
must admit the patient to activate alarms, automatic alarm printing
and the
ALARM HISTORY
.
WARNING
—After an interruption of the electric power supply, verify the alarm
processing is active and verify that no arrhythmia occurred.
CAUTION
—GE recommends that monitors without arrhythmia detection
monitoring patients at risk of asystole should define the HR alarm as
CRISIS
. If the HR alarm is set to
WARNING
, the alarm clears after
three minutes of continued asystole.
NOTE
Always remove all patient cables from the monitor prior to discharging a patient
to clear all alarm data from the monitor.
Control Settings
To adjust the control settings for the patient admitted to the monitor, select
MORE
MENUS
>
ALARM CONTROLS
. Then select any of the following options and make
the desired changes.
NOTE
All changes are temporary and revert to the factory or custom defaults when the
patient is discharged. To permanently change the these settings, refer to
Summary of Contents for Dash 3000
Page 11: ...2000966 386D Dash 3000 4000 5000 ix Symbols F 8 ...
Page 12: ...x Dash 3000 4000 5000 2000966 386D ...
Page 15: ...2000966 386D Dash 3000 4000 5000 1 1 1 Introduction ...
Page 33: ...2000966 386D Dash 3000 4000 5000 2 1 2 Equipment Overview ...
Page 48: ...2 16 Dash 3000 4000 5000 2000966 386D Equipment Overview ...
Page 49: ...2000966 386D Dash 3000 4000 5000 3 1 3 Monitor Setup ...
Page 61: ...2000966 386D Dash 3000 4000 5000 4 1 4 Alarms ...
Page 69: ...2000966 386D Dash 3000 4000 5000 5 1 5 Managing Patients ...
Page 81: ...2000966 386D Dash 3000 4000 5000 6 1 6 Patient Data ...
Page 92: ...6 12 Dash 3000 4000 5000 2000966 386D Patient Data ...
Page 93: ...2000966 386D Dash 3000 4000 5000 7 1 7 Printing ...
Page 97: ...2000966 386D Dash 3000 4000 5000 8 1 8 Monitoring ECG ...
Page 116: ...8 20 Dash 3000 4000 5000 2000966 386D Monitoring ECG ...
Page 117: ...2000966 386D Dash 3000 4000 5000 9 1 9 Monitoring Invasive Pressures ...
Page 125: ...2000966 386D Dash 3000 4000 5000 10 1 10 Monitoring NBP ...
Page 135: ...2000966 386D Dash 3000 4000 5000 11 1 11 Monitoring SpO2 ...
Page 149: ...2000966 386D Dash 3000 4000 5000 12 1 12 Measuring Cardiac Output ...
Page 155: ...2000966 386D Dash 3000 4000 5000 13 1 13 Monitoring Respiration ...
Page 159: ...2000966 386D Dash 3000 4000 5000 14 1 14 Monitoring Temperature ...
Page 162: ...14 4 Dash 3000 4000 5000 2000966 386D Monitoring Temperature ...
Page 163: ...2000966 386D Dash 3000 4000 5000 15 1 15 Monitoring End Tidal CO2 ...
Page 170: ...15 8 Dash 3000 4000 5000 2000966 386D Monitoring End Tidal CO2 ...
Page 171: ...2000966 386D Dash 3000 4000 5000 16 1 16 Monitoring Anesthesia Gases ...
Page 179: ...2000966 386D Dash 3000 4000 5000 17 1 17 Monitoring Impedance Cardiography ICG ...
Page 185: ...2000966 386D Dash 3000 4000 5000 18 1 18 Monitoring Bispectral Index BIS ...
Page 193: ...2000966 386D Dash 3000 4000 5000 A 1 A Maintenance ...
Page 210: ...A 18 Dash 3000 4000 5000 2000966 386D Maintenance ...
Page 211: ...2000966 386D Dash 3000 4000 5000 B 1 B Software Packages ...
Page 214: ...B 4 Dash 3000 4000 5000 2000966 386D Software Packages ...
Page 215: ...2000966 386D Dash 3000 4000 5000 C 1 C Factory Defaults ...
Page 227: ...2000966 386D Dash 3000 4000 5000 D 1 D Custom Defaults ...
Page 236: ...D 10 Dash 3000 4000 5000 2000966 386D Custom Defaults ...
Page 237: ...2000966 386D Dash 3000 4000 5000 E 1 E Interfacing with Non GE Devices ...
Page 255: ...2000966 386D Dash 3000 4000 5000 F 1 F Abbreviations and Symbols ...
Page 268: ...Index 6 Dash 3000 4000 5000 2000966 386D ...
Page 269: ......