Revision C
250cx Series Maternal/Fetal Monitor
15-3
2036946-001
Maintenance: Cleaning
Cleaning
General care and cleaning are required for the 250cx Series Monitor and its
accessories. If an accessory is not listed, consult the manufacturer’s instructions.
CAUTION
Unplug the monitor from the AC power source and detach all
accessories from the monitor. Do not immerse accessories in any
liquid. Do not use abrasive cloth or cleaners on monitor or
accessories.
Monitor Exterior
1.
The exterior surfaces of the equipment may be cleaned with a dampened, lint-
free cloth. Use one of the following approved solutions:
Commercial diluted bleach solution
Mild soap (diluted)
Commercial diluted ammonia solution
NOTE
: Always dilute cleaning solutions per manufacturers’
recommendations.
2.
Wipe off cleaning solutions with a clean dry cloth.
3.
Do not use a cleaning substance containing wax.
4.
Do not pour or spray water or any cleaning solution on the equipment or permit
fluids to run behind switches, into the connectors, into the recorder, or into any
ventilation openings in the equipment.
5.
Do not use the following cleaning agents:
Abrasive cleaners or solvents of any kind
Acetone
Ketone
Alcohol-based cleaning agents or
Betadine
CAUTION
Failure to follow these rules may melt, distort, or dull the finish of
the case, blur lettering on the labels, or cause equipment failures.
Cleaning products known to cause the types of problems
mentioned above include, but are not limited to Sani-Cloth
Wipes*, Sani-Wipes*, and Ascepti Wipes*. These should be
avoided. Products containing active ingredients and solutions
similar to these products should also be avoided.
Summary of Contents for Corometrics 250cx Series
Page 2: ......
Page 17: ...Revision C 250cx Series Maternal Fetal Monitor 1 1 2036946 001 1 Safety...
Page 18: ...1 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Safety For your notes...
Page 28: ...1 12 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Safety Equipment Symbols...
Page 29: ...Revision C 250cx Series Maternal Fetal Monitor 2 1 2036946 001 2 Introduction...
Page 30: ...2 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Introduction For your notes...
Page 53: ...Revision C 250cx Series Maternal Fetal Monitor 4 1 2036946 001 4 Setup Procedures...
Page 54: ...4 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Setup Procedures For your notes...
Page 65: ...Revision C 250cx Series Maternal Fetal Monitor 5 1 2036946 001 5 Fetal Heart Rate Monitoring...
Page 77: ...Revision C 250cx Series Maternal Fetal Monitor 6 1 2036946 001 6 Uterine Activity Monitoring...
Page 117: ...Revision C 250cx Series Maternal Fetal Monitor 10 1 2036946 001 10 Alarms...
Page 118: ...10 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Alarms For your notes...
Page 127: ...Revision C 250cx Series Maternal Fetal Monitor 11 1 2036946 001 11 Recorder Modes...
Page 128: ...11 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Recorder Modes For your notes...
Page 153: ...Revision C 250cx Series Maternal Fetal Monitor 13 1 2036946 001 13 Heartbeat Coincidence...
Page 159: ...Revision C 250cx Series Maternal Fetal Monitor 14 1 2036946 001 14 Waveforms...
Page 160: ...14 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Waveforms For your notes...
Page 165: ...Revision C 250cx Series Maternal Fetal Monitor 15 1 2036946 001 15 Maintenance...
Page 166: ...15 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Maintenance For your notes...
Page 174: ...16 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Troubleshooting For your notes...
Page 200: ...A 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Factory Defaults For your notes...
Page 212: ...B 6 250cx Series Maternal Fetal Monitor Revision C 2036946 001...
Page 214: ...C 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Spectra Alerts For your notes...
Page 236: ...D 2 250cx Series Maternal Fetal Monitor Revision C 2036946 001 Frequently Asked Questions...
Page 249: ......