4-16
250 Series Maternal/Fetal Monitor
Revision D
2020551-001
Maintenance:
Electrical Safety Tests
Patient-to-Line Leakage for US2
1. Connect an US test body to the monitor’s front panel
US2
input. The US test
body should be an ultrasound transducer wrapped in aluminum foil.
2. Set the fixture’s
Leakage
switch to
Line
.
3. Turn ON the monitor using the monitor’s power switch.
4. Verify the following:
< 50
µ
A (100, 120, 220, 230, or 240 VAC monitors)
5. Record the patient-to-ground leakage for the following conditions:
Ground Continuity
Use an ohm meter to confirm continuity between the monitor’s AC connector inlet
ground pin and unit’s chassis.
Dielectric (Hi-Pot) Tests
CAUTION
POWER OFF—Turn off the monitor prior to performing any of
the hi-pot tests.
Patient–to–AC-Line Using DC Voltage for 1 Minute
Connect the monitor to the hi-pot tester as shown on in the figure below.
NOTE:
The hi-pot tester voltage is 5.656 kVdc.
Table 8. Patient-to-Line Leakage Conditions for US2
Neutral
Ground
Power (Polarity)
Closed
Closed
Normal
Closed
Closed
Reversed
Open
Closed
Normal
Open
Closed
Reversed
Summary of Contents for Corometrics 250 Series
Page 2: ......
Page 6: ...CE CE ii 0086 ...
Page 14: ...viii 250 Series Maternal Fetal Monitor Revision D 2020551 001 ...
Page 16: ...1 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 For your notes ...
Page 29: ...Revision D 250 Series Maternal Fetal Monitor 2 1 2020551 001 2 Equipment Overview ...
Page 30: ...2 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 For your notes ...
Page 55: ...3 1 250 Series Maternal Fetal Monitor Revision D 2020551 001 3 Installation ...
Page 56: ...3 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 For your notes ...
Page 81: ...Revision D 250 Series Maternal Fetal Monitor 4 1 2020551 001 4 Maintenance ...
Page 82: ...4 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 For your notes ...
Page 143: ...Revision D 250 Series Maternal Fetal Monitor 5 1 2020551 001 5 Troubleshooting ...
Page 144: ...5 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 For your notes ...
Page 196: ...6 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 For your notes ...
Page 239: ...Revision D 250 Series Maternal Fetal Monitor B 1 2020551 001 B Alarms Summary ...
Page 240: ...B 2 250 Series Maternal Fetal Monitor Revision D 2020551 001 Alarms Summary For your notes ...
Page 243: ...Revision D 250 Series Maternal Fetal Monitor C 1 2020551 001 C Electromagnetic Compatibility ...
Page 251: ......