4-12
170 Series Monitor
Revision D
2003023-001
Setup Procedures:
Customizing the Monitor
Table 4-2. Summary of User Setup Codes
Code
(UA Display)
Setting or Value
(Primary FHR Display)
Code
Description
1
FHR Alarms
0 = off (disabled)
1 = on (enabled)
2
FHR High Alarm Limit
140–210 (BPM, in increments of 5 BPM)
3
FHR Low Alarm Limit
50–140 (BPM, in increments of 5 BPM)
4
FHR Alarm Volume
2–10
10
Minutes (time setting)
0–59 (minutes)
11
Hours (time setting)
0–23 (hours)
12
Day of Month (date setting)
1–31 (day)
13
Month (date setting)
1–12 (month)
14
Year (date setting)
00–99 (year)