5-6
CIC Pro™
2026419-033E
Installation
Do not remove the bottom feet or operate with the bottom of the unit on a carpeted
surface.
If installed with the vent holes in the bottom surface to the side (standing on a side, or
vertical installation), a minimum of 1 centimeter (3/8 inch) clearance must be
provided between the vent holes and the adjacent surface.
Electrical grounding
Duplex power outlet
A properly grounded duplex power outlet is required for each CIC Pro center.
NOTE
It is the customer’s responsibility to ensure that this requirement is met.
Additional outlets may be required to accommodate connected peripheral equipment.
The power outlet must be installed in an approved junction box. Use only a three-
prong, polarized, hospital-grade power outlet to accept the three-prong polarized CIC
Pro center power plug.
Grounding system
WARNING
SHOCK HAZARD — The CIC Pro center and all peripheral
equipment must be adequately grounded or a shock hazard may
exist. Do not use plug adapters that defeat the grounding capability
of the three-prong power plug. An ungrounded electrical device
presents a potentially severe and dangerous shock hazard.
The grounding pin of all power outlets and all exposed metal parts (beds, radiators,
water pipes, etc.) in any patient area should be electrically connected together. This
common ground point should be connected to the nearest equipotential ground
through a bonded grounding system, or with a 10 AWG stranded copper grounding
cable.
The equipotential ground point should be as close to earth ground potential as
possible. If a bonded grounding system is not available, the ground pin of each power
outlet must be individually connected to a central grounding point. Do not jumper
from ground pin-to-ground pin of the outlets.
The grounding system must not carry current, such as a grounded neutral, since the
current flow will produce potential differences along the ground path. These potential
differences are a shock hazard source for equipment users and patients.
Do
not
use conduit as a ground conductor. Plastic (PVC) piping or fittings used in the
conduit runs can break the electrical connection to ground, resulting in potential
shock hazards.
The electrical grounding system should be connected to an earth ground. If this is not
possible, then a good ground reference, such as a metal water pipe, or an electrically-
conductive building component, should be used. It is more important that all
grounded objects in the patient area are at the same ground potential rather than at
earth ground potential.
Summary of Contents for CIC Pro
Page 12: ...x CIC Pro 2026419 033E ...
Page 13: ...2026419 033E CIC Pro 1 1 1 Introduction ...
Page 27: ...2026419 033E CIC Pro 2 1 2 Equipment overview ...
Page 53: ...2026419 033E CIC Pro 3 1 3 Licensing ...
Page 60: ...3 8 CIC Pro 2026419 033E Licensing ...
Page 61: ...2026419 033E CIC Pro 4 1 4 Service interfaces ...
Page 74: ...4 14 CIC Pro 2026419 033E Service interfaces ...
Page 75: ...2026419 033E CIC Pro 5 1 5 Installation ...
Page 94: ...5 20 CIC Pro 2026419 033E Installation ...
Page 95: ...2026419 033E CIC Pro 6 1 6 Configuration ...
Page 187: ...2026419 033E CIC Pro 7 1 7 Checkout procedures ...
Page 208: ...7 22 CIC Pro 2026419 033E Checkout procedures ...
Page 209: ...2026419 033E CIC Pro 8 1 8 Troubleshooting ...
Page 241: ...2026419 033E CIC Pro 9 1 9 Field replaceable units FRUs ...
Page 243: ...Field replaceable units FRUs 2026419 033E CIC Pro 9 3 Hard drive Flash drive ...
Page 244: ...9 4 CIC Pro 2026419 033E Field replaceable units FRUs Rack mount ...
Page 245: ...Field replaceable units FRUs 2026419 033E CIC Pro 9 5 Mounting options ...
Page 265: ...2026419 033E CIC Pro 10 1 10 Preventive maintenance ...
Page 279: ...2026419 033E CIC Pro 11 1 11 Reload software ...
Page 288: ...11 10 CIC Pro 2026419 033E Reload software ...
Page 289: ...2026419 033E CIC Pro 12 1 12 Upgrade software ...
Page 298: ...12 10 CIC Pro 2026419 033E Upgrade software ...
Page 299: ...2026419 033E CIC Pro A 1 A Electromagnetic compatibility ...
Page 305: ...2026419 033E CIC Pro B 1 B Backup and restore backed up data modules ...
Page 309: ...Device compatibility 2026419 033E CIC Pro C 1 C Device compatibility ...
Page 312: ...C 4 CIC Pro 2026419 033E Device compatibility ...
Page 313: ......