Field replaceable units (FRUs)
2026419-033E
CIC Pro™
9-15
CAUTION
NETWORK DEVICE TIME SYNCHRONIZATION — When
adding a new device (e.g., CIC Pro center) to the CARESCAPE
Network, the existing devices on the CARESCAPE Network will
synchronize to the new device’s time. To prevent potential time
synchronization issues, you should set the new device’s time to be as
close as possible to the time (within a few seconds) used by the
existing GE devices on the CARESCAPE Network.
Check the status of installed licenses on page 7-4
Check access to all other care units on page 7-7
.
Check the pressures unit-of-measure on page 7-8
(for Chinese applications
only).
Verify that the Remote Service agent is enabled.
Remote Service Agent on page 6-18.
Check the laser printer connectivity.
See Check status of installed printers on
Check the digital writer printer connectivity.
Check MultiKM (Multimouse) operation on page 7-8
Check full disclosure report printing on page 7-9
Check read and write integrity of hard drive and solid state flash drive on
page 7-10
.
Check audio component operation on page 7-16
.
Check integrity of system files on page 7-20
.
Check operation of the secondary display on page 7-4
.
.
Check BIOS information on page 7-13
Check drive operation information (Flash drive and hard drive) on page 7-
14
.
Check operation of the Watchdog countdown function on page 7-18
.
Check video function and status of video card and drivers on page 7-19
.
Check internal hardware temperature and voltage status on page 7-12
Check system resources.
See Environment Monitor messages on page 8-19.
Verify that the power LED is on after the unit is powered up.
NOTE
Do not re-image the Flash drive or hard drive after replacing the CPU PCB.
Replacing the CPU battery
Once the battery is replaced, the system loses its time and the time must be reset
before connecting the CIC Pro center to the existing CARESCAPE Network.
Summary of Contents for CIC Pro
Page 12: ...x CIC Pro 2026419 033E ...
Page 13: ...2026419 033E CIC Pro 1 1 1 Introduction ...
Page 27: ...2026419 033E CIC Pro 2 1 2 Equipment overview ...
Page 53: ...2026419 033E CIC Pro 3 1 3 Licensing ...
Page 60: ...3 8 CIC Pro 2026419 033E Licensing ...
Page 61: ...2026419 033E CIC Pro 4 1 4 Service interfaces ...
Page 74: ...4 14 CIC Pro 2026419 033E Service interfaces ...
Page 75: ...2026419 033E CIC Pro 5 1 5 Installation ...
Page 94: ...5 20 CIC Pro 2026419 033E Installation ...
Page 95: ...2026419 033E CIC Pro 6 1 6 Configuration ...
Page 187: ...2026419 033E CIC Pro 7 1 7 Checkout procedures ...
Page 208: ...7 22 CIC Pro 2026419 033E Checkout procedures ...
Page 209: ...2026419 033E CIC Pro 8 1 8 Troubleshooting ...
Page 241: ...2026419 033E CIC Pro 9 1 9 Field replaceable units FRUs ...
Page 243: ...Field replaceable units FRUs 2026419 033E CIC Pro 9 3 Hard drive Flash drive ...
Page 244: ...9 4 CIC Pro 2026419 033E Field replaceable units FRUs Rack mount ...
Page 245: ...Field replaceable units FRUs 2026419 033E CIC Pro 9 5 Mounting options ...
Page 265: ...2026419 033E CIC Pro 10 1 10 Preventive maintenance ...
Page 279: ...2026419 033E CIC Pro 11 1 11 Reload software ...
Page 288: ...11 10 CIC Pro 2026419 033E Reload software ...
Page 289: ...2026419 033E CIC Pro 12 1 12 Upgrade software ...
Page 298: ...12 10 CIC Pro 2026419 033E Upgrade software ...
Page 299: ...2026419 033E CIC Pro A 1 A Electromagnetic compatibility ...
Page 305: ...2026419 033E CIC Pro B 1 B Backup and restore backed up data modules ...
Page 309: ...Device compatibility 2026419 033E CIC Pro C 1 C Device compatibility ...
Page 312: ...C 4 CIC Pro 2026419 033E Device compatibility ...
Page 313: ......