EEG and auditory evoked potentials
Using the EEG measurement
Starting the EEG measurement
EEG measurement starts automatically.
1.
Observe the results of the electrode check.
2.
Reconnect the electrodes if the impedance is too high.
Selecting your own EEG montage
If you are using your own montage, you will have to select it from the list.
Preconfigured montages will be automatically recognized.
1.
Select the EEG parameter window.
2.
Select
EEG
tab >
Montage
.
3.
Select the montage from the
Select Montage
list.
NOTE
Since the monitor sets lead positions automatically according
to the identification pin, make sure that you use preconfigured
lead positions. If you change them, always update the
montage settings according to the instructions.
Selecting the EEG scale
This setting defines the scale for EEG parameter windows, waveform and spectral
displays
1.
Select the EEG parameter window.
2.
Select
EEG
tab >
Setup
.
3.
Select a value from the
Scale µV
list.
Selecting the EEG sweep speed
This setting determines the drawing speed for the EEG waveform.
1.
Select the EEG parameter window.
2.
Select
EEG
tab >
Setup
.
3.
Select a value from the
EEG Sweep Speed
list.
The smaller the value, the slower the sweep speed.
NOTE
This setting is available in BIS, EEG and Entropy setups.
Regardless of where you change it, it will affect all three
parameters.
Selecting EEG numeric parameters
EEG numeric trends are available in the ED, ICU, and NICU software packages.
1.
Select the EEG parameter window.
2.
Select
EEG
tab >
Setup
.
3.
Select a parameter from the
Numeric 1
list.
2062971-001
CARESCAPE Modular Monitors
379
Summary of Contents for CARESCAPE
Page 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Page 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Page 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Page 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Page 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Page 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Page 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Page 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Page 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Page 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Page 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Page 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Page 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Page 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Page 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Page 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Page 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Page 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Page 547: ...content ...