EEG and auditory evoked potentials
CAUTION
EEG measurement is inherently very sensitive. Radiated
electromagnetic fields may cause erroneous measurements
at various frequencies. Do not use electrical radiating
equipment close to the EEG measurement module. Details
regarding radiated field strengths are given in the technical
specifications.
EEG indications for use
The EEG module, E-EEG, and the headbox, N-EEG, and accessories are indicated for
the monitoring of electroencephalograph (EEG), frontal electromyography (FEMG), and
auditory evoked potentials (AEP) of all hospital patients.
The device is indicated for use by qualified medical personnel only.
EEG measurement limitations
●
This measurement is not available in the NICU software package.
●
E-modules are not suitable for use with neonatal patients.
●
External radiating devices may disturb the measurement.
EEG points to note
●
EEG is a very small signal and prone to artifacts such as movement, shivering,
sweating, eye movement, frontal EMG, external electrical interference (other
electrical devices), and internal (ECG) electrical interference.
●
Good electrode contacts are imperative for high-quality EEG measurement.
●
Keep electrical devices as far away from the patient as possible.
●
Use short cables, or braid the long ones.
●
Keep the headbox as close to the patient's head as possible.
●
EEG findings are usually sensitive but not specific to the underlying cause.
●
EEG findings should be assessed in conjuction with:
■
other monitor data and clinical assessment
■
neurological status of the patient
■
factors that influence EEG (like blood flow, neurological pathologies, underlying
illnesses, and temperature).
●
A cloth bandage over cup electrodes and around the patient's head may enhance
good contact.
374
CARESCAPE Modular Monitors
2062971-001
Summary of Contents for CARESCAPE
Page 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Page 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Page 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Page 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Page 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Page 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Page 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Page 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Page 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Page 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Page 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Page 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Page 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Page 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Page 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Page 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Page 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Page 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Page 547: ...content ...