Mixed venous oxygen saturation (SvO₂)
SvO
2
measurement description
The SvO
2
value is measured continuously by spectrophotometry. The algorithm
consists of five different parts:
●
Initialization: When connected, a number of start-up procedures are performed
prior to normal operation. These procedures include transfer of calibration factors
and initialization of LED currents.
●
Calibration
●
Signal processing and SvO
2
calculation: Light of various wavelengths (red 660 nm
and infrared 810 nm) is transmitted to the blood through a single plastic optical
fiber in the oximetry catheter, and reflected back through a separate optical fiber
to a photodetector. The light is electrically transmitted and analyzed. From the
amount of reflected light it is possible to measure the amount of light absorbed
by hemoglobin and oxyhemoglobin, resulting in the SvO
2
value. The SvO
2
value
is displayed as a percentage.
●
Automatic gain control: The intensity of the red and infrared signals can be
amplified by four different gains. The gain is selected automatically to achieve
optimal signal levels.
●
Signal quality
SvO
2
troubleshooting
Problem
Solution
SvO
2
levels are too high
●
Position the catheter correctly.
●
Calibrate in vivo.
In vivo calibration fails
●
Check the connections.
●
Check that the optical cables have no sharp bends.
●
If the in vivo calibration fails again, replace the catheter and/or the
optical module, and repeat the procedure.
2062971-001
CARESCAPE Modular Monitors
279
Summary of Contents for CARESCAPE
Page 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Page 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Page 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Page 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Page 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Page 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Page 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Page 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Page 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Page 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Page 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Page 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Page 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Page 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Page 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Page 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Page 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Page 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Page 547: ...content ...