Invasive pressures
Invasive pressures compatibility limitations
TRAM and Tram-Rac modules with the B850 only.
For detailed information regarding module, monitor, and accessory compatibility, see
the CARESCAPE Modular Monitors Supplemental Information Manual.
Invasive pressure safety precautions
Invasive pressure warnings
WARNING
All invasive procedures involve risks to the patient. Use aseptic
technique. Incorrect use of the catheter can lead to vessel
perforation. Follow catheter manufacturer's instructions.
WARNING
Make sure that no part of the patient connections touches
any electrically conductive material including earth.
WARNING
Mechanical shock to an invasive blood pressure transducer
may cause severe shifts in the zero balance and calibration,
and cause erroneous readings.
WARNING
Repositioning the patient after a completed zeroing procedure
may cause incorrect measurement values.
WARNING
When connecting PDM or TRAM, the loaded IP labels may
affect the channel labeling of other already connected
channels, and consequently also the alarm limits.
Invasive pressure measurement limitations
●
E-modules used for this measurement are not suitable for use with neonatal
patients.
Invasive pressure points to note
●
Depending on the invasive pressure module used, not all invasive pressure
measurements and settings are available to view or change.
2062971-001
CARESCAPE Modular Monitors
235
Summary of Contents for CARESCAPE
Page 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Page 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Page 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Page 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Page 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Page 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Page 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Page 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Page 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Page 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Page 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Page 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Page 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Page 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Page 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Page 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Page 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Page 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Page 547: ...content ...