Non-invasive blood pressure
NIBP troubleshooting
Problem
Solution
NIBP measurement does not work or
the values seem unstable.
●
Check that the cuff tubing is not bent, stretched, compressed,
or loose.
●
Check the cuff position and cuff tube connection.
●
Prevent motion artifact.
●
Use NIBP cuffs of correct size.
Why does the mean value display while
the associated systolic and diastolic
values display as - - -?
●
Assess the patient and perform a visual inspection of the
equipment to ensure system integrity.
The following conditions may cause the mean value to display in
the NIBP parameter window while the associated systolic and
diastolic values display as - - -:
●
Very low systolic and diastolic amplitude fluctuations (e.g.,
patient in shock).
●
Very small difference between the mean and systolic pressure
or the mean and diastolic pressure.
●
Loss of system integrity (e.g., loose connections or worn parts)
Why is the monitor re-inflating the cuff
automatically?
The cuff target pressure must be higher than the patient’s systolic
pressure to obtain an accurate systolic and diastolic measurement.
If a systolic blood pressure cannot be found, the monitor searches
for a systolic reading by re-inflating the cuff at a higher pressure.
This systolic search may occur once per NIBP determination cycle.
During a systolic search, the maximum cuff inflation pressure
will not exceed the normal pressure range of the cuff. For more
information, refer to the technical specifications.
For PDM and PSM, the monitor may be taking a control
measurement. If the measured NIBP value exceeds the alarm
limits, a single low priority alarm sounds and a new measurement
is automatically taken. If the new value (the control measurement)
also exceeds the alarm limits the alarm priority escalates to
medium. In
Manual
mode and
STAT
mode there are at least
four seconds between the first measurement and the control
measurement for
Adult
and
Child
cuffs, eight seconds for
Infant
cuffs. In
Auto
mode there are at least 30 seconds between the first
measurement and the control measurement.
For PSM, if
Weak pulsation
or
Artifacts
is detected, the module
repeats the measurement up to three times. Assess the patient
and perform a visual inspection of the equipment to ensure system
integrity.
For PSM, if cuff occlusion is detected during inflation the module
might repeat the measurement. Perform a visual inspection of the
equipment to ensure system integrity.
2062971-001
CARESCAPE Modular Monitors
233
Summary of Contents for CARESCAPE
Page 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Page 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Page 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Page 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Page 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Page 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Page 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Page 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Page 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Page 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Page 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Page 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Page 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Page 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Page 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Page 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Page 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Page 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Page 547: ...content ...