Non-invasive blood pressure
●
x = Cuff pulsation waveform
●
y = Cuff pressure
1.
Cuff deflation
2.
Systolic pressure (ratio of maximum amplitude)
3.
Mean arterial pressure (maximum pulsation amplitude)
4.
Diastolic pressure (ratio of maximum amplitude)
5.
Cuff pulsations (each pulsation represents one heart beat)
6.
Amplitude (changes based on cuff pressure)
7.
Resulting waveform
Systolic and diastolic determinations are based on a mathematical calculation within
the algorithm. The deflation mode is heart rate dependent, it is typically longer with
heart rates that are slow and/or irregular.
This patented process of finding two matched pulsations of relatively equal amplitude
and frequency at each step rejects artifact due to patient movement or other
deviations from ideal conditions (e.g., cuff disturbances) and greatly enhances the
overall accuracy of the monitor.
NOTE
NIBP values are based on the oscillometric method of
non-invasive blood pressure measurement taken with a cuff
on the arm of adult and child patients, and a cuff on the
calf of infants. The values correspond to comparisons with
intra-arterial values within ANSI/AAMI SP10 and IEC particular
standards for accuracy (a mean difference of ± 5 mmHg, and
a standard deviation of < 8 mmHg).
TRAM modules’ NIBP technology
TRAM modules use the DINAMAP Classic algorithm. This algorithm measures the
mean arterial pressure and then determines the systolic and diastolic values. The
algorithm examines pulses from the cuff and waits for two sequential pulses equal
in both frequency and amplitude. It will not deflate the cuff until it has good data.
Artifact and false pulses are rejected.
230
CARESCAPE Modular Monitors
2062971-001
Summary of Contents for CARESCAPE
Page 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Page 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Page 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Page 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Page 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Page 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Page 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Page 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Page 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Page 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Page 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Page 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Page 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Page 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Page 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Page 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Page 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Page 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Page 547: ...content ...