Equipment Overview
2002938-002E
Cardiolab® II Plus Amplifier Manual
2-9
Internal Assembly
Board Identifications
Control Board (PN: 2002344-001)
Provides communications between the amplifier and the CPU.
Transmits all real-time data.
Receives all amplifier setup information from the CPU and sends
commands to other boards
Table 6. Back View - CardioLab II Plus Amplifier
Name
Description
FDDI Connection to
Computer
This connector will accept a FDDI type fiber optic cable. The
fiber optic cable allows for two-way communication between
the amplifier and the CardioLab system.
Power Input
The GE-supplied power cord connects to the Power Input.
Equipotentiality
Terminal
The Equipotentiality Terminal is connected to earth ground.
Summary of Contents for Cardiolab II Plus
Page 1: ...Service Manual 2002938 002 Revision E Cardiolab II Plus Amplifier Manual ...
Page 8: ...i iv Cardiolab II Plus Amplifier Manual 2002938 002E ...
Page 9: ...2002938 002E Cardiolab II Plus Amplifier Manual 1 1 1 Introduction ...
Page 26: ...1 18 Cardiolab II Plus Amplifier Manual 2002938 002E Introduction ...
Page 27: ...2002938 002E Cardiolab II Plus Amplifier Manual 2 1 2 Equipment Overview ...
Page 36: ...2 10 Cardiolab II Plus Amplifier Manual 2002938 002E Equipment Overview ...
Page 37: ...Equipment Overview 2002938 002E Cardiolab II Plus Amplifier Manual 2 11 ...
Page 47: ...2002938 002E Cardiolab II Plus Amplifier Manual 3 1 3 Operation ...
Page 65: ...2002938 002E Cardiolab II Plus Amplifier Manual B 1 B Appendix B Technical Information ...
Page 67: ...Appendix B Technical Information 2002938 002E Cardiolab II Plus Amplifier Manual B 3 ...
Page 68: ...B 4 Cardiolab II Plus Amplifier Manual 2002938 002E Appendix B Technical Information ...
Page 69: ......