Chapter 2. Installation
28
PanaFlow™ LC User’s Manual
2.11.2.1 Transducer Wiring (ATEX installations)
1.
Thread one end of armored cable into the meter head and the other into the junction box on the back
of the transducer (see Figure 1 on page 1 for reference).
2.
Terminate connections of the transducer end of each cable to each of your clamp on transducers.
3.
At the meter end of the cable secure the red wire of channel one downstream RG62 cable to the CH1
(red box in Figure 41) DN terminal of the XMT1000 backplane.
4.
Secure Black wire of channel one downstream RG62 cable to the CH1 (red box in Figure 41) RTN
terminal directly below the DN terminal of the XMT1000 backplane.
5.
Secure red wire of channel one upstream RG62 cable to the CH1 (red box in Figure 41) UP terminal of
the XMT1000 backplane.
6.
Secure Black wire of channel one upstream RG62 cable to the CH1 (red box in Figure 41) RTN
terminal directly above the UP terminal of the XMT1000 backplane.
Figure 41: XMT1000, Channel Back Plane
Summary of Contents for Baker Hughes PanaFlow LC XMT1000
Page 1: ...Flow 910 327 Rev A Sep 2019 bhge com PanaFlow LC User s Manual ...
Page 2: ......
Page 4: ...ii no content intended for this page ...
Page 10: ...Contents viii PanaFlow LC User s Manual ...
Page 14: ...Preface xii PanaFlow LC User s Manual no content intended for this page ...
Page 142: ...Appendix E CE Mark Compliance 128 PanaFlow LC User s Manual no content intended for this page ...
Page 182: ...Warranty 168 PanaFlow LC User s Manual no content intended for this page ...
Page 189: ......