PanaFlow™ LC User’s Manual
163
Appendix H. Foundation Fieldbus Communication
H.12 Simulation Mode
Simulation mode
allows the user to test the FF implementation without the instrument providing real data.
The meter PCB is shipped with simulation mode disabled. To enable simulation mode, complete the
following steps:
1.
Remove the PCB from the meter.
2.
Locate the jumper
P5
below).
3.
Move the
P5
jumper to the left one place (pins 2 & 3) to enable simulation mode.
4.
Re-install the PCB into meter
5.
Verify that the
Block
error field in the
Resource Block
shows
SimulationActive
status.
Figure 141: P5 on XMT1000 PCB
CAUTION!
To prevent damage to the electronic components, always use
ESD
protection whenever
handling printed circuit boards.
Jumper P5