14
2062665-001 G
PCD-01: PCD-01 description
OBX-4
Condition predicate:
This field is used to distinguish between multiple OBX segments by providing an
unambiguous mapping from observation contained in the OBX segment to the
IEEE 11073 containment tree for the Medical Device System sourcing of the
observation. For device related data, this field groups devices hierarchically.
For metric related data, this field associates metrics to devices hierarchically and
is used as a sequence number within the hierarchical representation. The dotted
notation provided for in the object diagram is:
<MDS>.<VMD>.<Channel>.<Metric> for the PCD-01 transaction, where the
Metric is a sequence number.
NOTE:
Numeric and Enumeration are not used for PCD-01 transactions as
shown in the object diagram.
Physiological Monitor
Containment tree
Physiological monitors are comprised of a number of different Virtual Medical
Devices (VMDs) as indicated in the following example containment tree:
The first OBX following the OBR segment is always set to:
OBX|1||69965^MDC_DEV_MON_PHYSIO_MULTI_PARAM_MDS^MDC|1
.0.0.0|||||||X
This is the standard device reference used for a monitor or monitor system.
The second OBX following OBX(1) is a Virtual Medical Device (VMD) Header, e.g.,
OBX|10||69798^MDC_DEV_ECG_VMD^MDC|1.5.0.0|||||||X
Physiological Monitor Containment tree
Reference IDs
MDS: Physiological Monitor
MDC_DEV_METER_PHYSIO_MULTI_PARAM_MDS (1::4301)
VMD: Blood Pressure
MDC_DEV_METER_PRESS_BLD_VMD (1::4318)
Channel: Invasive BP
MDC_DEV_METER_PRESS_BLD_CHAN (1::4319)
Channel: Non-Invasive BP
MDC_DEV_PRESS_BLD_NONINV_CHAN (1::5151)
VMD: Temperature
MDC_DEV_METER_TEMP_VMD (1::4366)
VMD: Pulse Oximeter
MDC_DEV_ANALY_SAT_O2_VMD (1::4106)
VMD: ECG Monitor
MDC_DEV_ECG_VMD (1::4262)
VMD: Entropy
MDC_DEV_EEG_ENTROPY_VMD
Summary of Contents for B40
Page 5: ...2062665 001 G 3 B Abbreviations 45 Abbreviations 46...
Page 6: ...4 2062665 001 G...
Page 7: ...2062665 001 G 1 Introduction 1 Introduction...
Page 11: ...2062665 001 G 5 HL7 Interface Service Information 2 HL7 Interface...
Page 18: ...12 2062665 001 G PCD 01 HIS Disconnection or Network Outage 4 PCD 01...
Page 23: ...2062665 001 G 17 Sample Messages PCD 01 description 5 Sample Messages...
Page 43: ...2062665 001 G 37 HL7 Data Segment Definitions Data Segments A Supported parameters...
Page 51: ...2062665 001 G 45 HL7 Data Segment Definitions Supported parameters B Abbreviations...
Page 54: ...48 2062665 001 G HL7 Data Segment Definitions Abbreviations...
Page 55: ......