B30 Patient Monitor
4-14
Enabling or disabling alarm silencing
With this selection, you can determine whether the audible alarms can be turned off or not.
1.
Press the
Monitor Setup
key.
2.
Select
Install/Service
and enter the password.
3.
Select
Installation - Alarm Options
.
4.
Select
Show
Audio ON/OFF
and
YES
to enable alarm silencing or
NO
to disable it.
Disabling the alarm light
By default, the
Alarm Light
has been enabled. The light flashes red or yellow according to the
currently active alarm. If several alarms are active at the same time, the color follows the
highest priority alarm.
To disable the alarm light:
1.
Press the
Monitor Setup
key.
2.
Select
Install/Service
and enter the password.
3.
Select
Installation
-
Alarm Options
.
4.
Select
Alarm Light
-
NO
.
Monitor
Setup
Monitor
Setup
Summary of Contents for b30
Page 2: ......
Page 8: ......
Page 10: ...Table of contents ii...
Page 12: ...Table of contents ii...
Page 16: ...B30 Patient Monitor 4...
Page 18: ...Table of contents ii...
Page 26: ...Table of contents ii...
Page 62: ...B30 Patient Monitor 2 36...
Page 64: ...Table of contents ii...
Page 72: ...Table of contents ii...
Page 90: ...Table of contents ii...
Page 106: ...Table of contents ii...
Page 118: ...B30 Patient Monitor 6 12...
Page 120: ...Table of contents ii...
Page 124: ...B30 Patient Monitor 7 4...
Page 126: ...Table of contents ii...
Page 136: ...Table of contents ii...
Page 144: ...B30 Patient Monitor 9 8...
Page 146: ...Table of contents ii...
Page 156: ...Table of contents ii...
Page 176: ...Table of contents ii...
Page 184: ...B30 Patient Monitor 12 8...
Page 186: ...Table of contents ii...
Page 192: ...Table of contents ii...
Page 202: ...Table of contents ii...
Page 212: ...B30 Patient Monitor 15 10...
Page 214: ...Table of contents ii...
Page 224: ...B30 Patient Monitor 16 10...
Page 226: ...Table of contents ii...
Page 236: ...B30 Patient Monitor 17 10...
Page 241: ......