Inspiratory Breathing resistance in bag mode*
l/min
kPa
cmH2O
5
0.06
0.6
30
0.21
2.1
60
0.49
5.0
*Values include patient circuit tubing and wye-piece 0.049 kPa (0.5
cmH2O) at 60 l/min. Patient circuit tubing and breathing system
configurations affect resistance.
Expiratory Breathing resistance in bag mode*
l/min
kPa
cmH2O
5
0.05
0.5
30
0.15
1.5
60
0.37
3.8
EZchange canister system and condenser, absorber mode
5
0.06
0.6
30
0.19
1.9
60
0.44
4.5
EZchange canister system and condenser, canister removed
5
0.06
0.6
30
0.19
1.9
60
0.44
4.5
*Values include patient circuit tubing and wye-piece 0.03 kPa (0.3 cmH2O)
at 60 l/min. Patient circuit tubing and breathing system configurations
affect resistance.
Pressure flow data (APL valve completely open)
Flow (l/min)
Flow (l/s)
APL pressure cmH2O
3
0.05
0.93
30
0.51
1.70
60
1.0
2.44
70
1.17
3.30
Gas scavenging
All scavenging
Positive pressure relief
10 cmH2O
Aisys CS²
11-14
2067226-001
Summary of Contents for auisys cs2
Page 1: ...Aisys CS User s Reference Manual Software Revision 10 X ...
Page 32: ...1 20 2067226 001 ...
Page 60: ...3 Set the alarm limits to clinically appropriate settings Aisys CS 3 12 2067226 001 ...
Page 87: ...3 Push the drain button to empty any water in the condenser 3 Operation 2067226 001 3 39 ...
Page 90: ...3 42 2067226 001 ...
Page 94: ...4 4 2067226 001 ...
Page 132: ...7 20 2067226 001 ...
Page 164: ...9 8 2067226 001 ...
Page 178: ...System pneumatic circuit AB 75 270 Aisys CS 11 2 2067226 001 ...
Page 230: ...4 Select Back to return to the Gas Usage menu Aisys CS 12 4 2067226 001 ...
Page 262: ...13 20 2067226 001 ...
Page 268: ...Aisys CS User s Reference Manual English 2067226 001 06 13 C 18 05 02 Printed in USA ...