background image

2

28-9266-70 AA

Sightglass cover assembly instruction

Assembly mount procedure – 6 mm and 1/2 inch air traps

1

Disconnect the air trap and remove the bottom nuts

Remove the TC coupling (A) and evacuation hose from the 
top valve.

Unscrew and disconnect the cable contacts (B) from the 
level sensors.

Disconnect the air trap inlet and outlet from the system 
valves.

Loosen and remove four bottom nuts (E) and washers from 
the connecting rods that are fixed to the support (D).

Note:

The nut on the fifth connecting rod does not need to be 
removed.

Note:

The washers from the two front connecting rods will not 
be used when the cover is mounted.

Lift the air trap from the support (D) and move it to a 
suitable place to mount the cover assembly.

2

Pull out the connecting rods

Remove the bottom nuts (C) from the two front connecting 
rods.

Pull the connecting rods (F) out of the air trap. Remove the 
washers from the rods.

Note:

Do not remove the dome nuts. Ensure that the dome nuts 
still are securely fastened. If loose, the nuts should be 
fastened and locked into place using Loctite™ 243.

3

Mount the top sightglass cover part

Place the top cover (G) on the air trap, with the opening 
aligned with the sight glass.

Re-insert the front two connecting rods (F) through the top 
holes in the cover.

Insert a plain washer and a lock washer on each of the 
two front connecting rods underneath the air trap and 
mount the nuts (C).

Note:

The lock washer should be closest to the nut.

Tighten the nuts using a torque of 3 Nm (±0.3 Nm). Verify 
also that the other three nuts are tightened, using the 
same torque value.

Return the air trap to the support (D), aligning the bottom 
part of each connecting rod through the holes in the 
support.

Connect the air trap inlet and outlet to the system valves.

A

B

E

D

C

F

F

G

Reviews: