ÄKTApilot

Site Preparation Guide

Summary of Contents for AKTApilot

Page 1: ...ÄKTApilot Site Preparation Guide ...

Page 2: ...Table of Contents 3 1 Introduction 5 2 Site preparation overview 10 3 Site preparation checklist 2 ÄKTApilot Site Preparation Guide 29 1396 54 AA Table of Contents ...

Page 3: ...This Site Preparation Guide covers the ÄKTApilot system The ÄKTApilot system consists of the following components ÄKTApilot system separation unit Computer including a flat screen monitor UNICORN control system Feedback to GE Make sure to read the entire Site Preparation Guide and complete the Site preparation checklist in Chapter 3 Return the completed Site preparation checklist to the GE service...

Page 4: ...and tasks Coordinate the installation timing of the system with any domestic safety procedures and testings Order the required materials Schedule the installation and inform the relevant personnel of the instal lation date Laboratory personnel Primary users Review the safety information Make sure that all customer provided materials for the installation are present at the site Facilities personnel...

Page 5: ...irements must be fulfilled at the location where the ÄKTApilot system will be installed ÄKTApilot is intended for indoor use only The instrument units must not be installed in a dusty environment The instrument units must not be exposed to direct sunlight Space requirements Prepare a stable laboratory bench with a height of 1200 mm The following space re quirements must be fulfilled at the locatio...

Page 6: ...e to follow these when connecting them Note WARNING Access to power cord Do not block access to the power cord The power cord must always be easy to disconnect Dimensions and weight The following illustration shows the dimensions of the ÄKTApilot system 900 mm 540 mm 750 mm The weight of the ÄKTApilot system is 114 kg monitor arm and computer not included 6 ÄKTApilot Site Preparation Guide 29 1396...

Page 7: ...ents The installation site must comply with the following specifications Specification Parameter 4 C to 40 C Ambient temperature operation 20 to 95 non condensing Relative humidity 840 to 1060 mbar Atmospheric pressure Electrical requirements Specification Parameter 100 to 240 V AC Supply voltage 50 60 Hz Frequency 800 VA Max power 10 A 250 V or 16 A 125 V Fuse 4 Number of wall outlets One for the...

Page 8: ... Professional 32 bit or 64 bit Operating system UNICORN 5 or higher versions Control system Note The computer should be installed placed and used according to the instruc tions and specifications provided by the manufacturer of the computer A printer should be connected to the computer The computer should not be connected to a network before the installation User supplied accessories and consumabl...

Page 9: ...ation Item For example WFI MQ purified High quality water 1 in high quality water 0 4 in 1 M sodium chloride Acetone 20 Ethanol pH 4 00 200 ml pH 7 00 200 ml pH 10 00 200 ml Calibration solutions for pH Approximately 2 12 and 110 mS cm Conductivity solutions All solutions should be filtered and degassed and all solutions should have relevant certificates Note Other product specifications Specifica...

Page 10: ...equirement General When was the ÄKTApilot system received month year Will the GE service personnel be required to take site EHS Environment Health and Safety or secu rity training before doing the installation Assigned personnel Personnel are assigned as specified Does the personnel experience level with the ÄKTApilot system correspond to new user intermediate user advanced user Will the personnel...

Page 11: ...ter has been purchased and received Has an administrator account been creat ed and a logon password established for our use Will the computer be connected to a net work Is there on site IT support available if needed Are the computer requirements fulfilled User supplied equipment Personal Protective Equipment PPE as required e g gloves lab coats etc Printer connected to the system computer Calibra...

Page 12: ...inges etc User supplied consumables High quality water filtered 10 liters Ethanol 20 3 liters Acetone 1 in high quality water 10 liters 0 4 in 1 M sodium chloride 5 liters pH solutions pH 4 00 7 00 and 10 00 Conductivity solutions approximately 2 12 and 110 mS cm Additional information 12 ÄKTApilot Site Preparation Guide 29 1396 54 AA 3 Site preparation checklist ...

Page 13: ... dd mmm yyyy Company Customer name Company Customer representative Name Address Email Phone number Advanced Intermediate Beginner Customer experience level ÄKTApilot Site Preparation Guide 29 1396 54 AA 13 3 Site preparation checklist ...

Page 14: ...ale of the company within GE Healthcare which supplies them A copy of these terms and conditions is available on request Contact your local GE Healthcare repre sentative for the most current information GE Healthcare Europe GmbH Munzinger Strasse 5 D 79111 Freiburg Germany GE Healthcare UK Limited Amersham Place Little Chalfont Buckinghamshire HP7 9NA UK GE Healthcare Bio Sciences Corp 800 Centenn...