September 2007
32
HygroTrace Moisture Transmitter
Erase
Note:
Erase
will delete ALL calibration and user settings. The
HygroTrace
will be incapable of performing accurate
measurements until this data is restored.
Note:
This function is provided to assist GE Sensing service
personnel in maintaining the product.
To erase the EPROM, select
Erase
,
press
Enter
and the following
message appears at the top of the
screen for a few seconds.
Select
YES
or
NO
and press
Enter
, or
press
Cancel
. If you select
NO
or
press
Cancel
, the unit returns to the
previous screen. If you select
YES
,
the following message appears.
When the erasure is complete,
Param Storage
reappears. To return
to the
Service Menu
, press
Cancel
.
Heater Test
To perform a test of the heater, select
Heater Test...
, press
Enter
and a
screen similar to the following
appears:
Use the
arrow
keys to select the
appropriate temperature to test, and
press
Enter
. The temperature in the
upper right corner of the screen
should approximate the one selected.
To turn off the test, select
OFF
and press
Enter
. The temperature
in the upper right corner of the screen returns to its original value.
To resume normal operation, select
RESUME
and press
Enter
.
Press
Cancel
to return to the
Service Menu
.
+
+
+
+