4 - 22
Section 4-2 - Functional checks
PW/CW Doppler Mode Checks
For information on the Ultrasound system’s different modes, the Basic User Manual or User Guide
will familiarize you with image optimization for
B-Mode
,
M-Mode
,
Color Flow
, and
Doppler
.
Doppler is used to measure velocity (most often in blood). Doppler mode can be done with a special
pencil probe or with an ordinary probe. By using an ordinary probe, you can first bring up a B-Mode
picture for navigation purpose and then add Doppler.
For information, refer to the appropriate version of the LOGIQ E10 Basic User Manual, Chapter 5,
or the appropriate LOGIQ E10 Release Notes.
Preparations
1.) Connect one of the probes to the scanner.
-
See:
Section 3-4-6 "Connecting Probes" on page 3-19
for info about connecting the probes.
-
For available probes, see: Section 9, Probes
.
2.) Turn ON the scanner
3.) The 2D Mode window is displayed (default mode).
4.) If needed, adjust the Display’s Brightness and Contrast setting.
5.) Press
PW
or
CW
to start Pulsed Wave Doppler (PW) or Continuous Wave Doppler (CW).
6.) Use the Trackball to select the Area of Interest (Sample Volume) in PW or direction of interest
in CW.
Adjust the PW/CW Doppler Mode controls
Adjust the
Active mode gain
to set the gain in the spectral Doppler area.
-
Adjust
Low velocity reject
to reduce unwanted low velocity blood flow and tissue movement.
-
In PW mode, adjust
Sample volume
to low setting for better resolution, or higher setting to
more easily locate the disturbed flows.
-
Adjust the
Compress
setting to balance the effect of stronger and weaker echoes and obtain
the desired intensity display.
-
Adjust
Frequency
to optimize flow display. Higher setting will improve resolution and the
lower setting will increase the depth penetration.
-
Adjust
Frame rate
to a higher setting to improve motion detection, or to a lower setting to
improve resolution.
NOTE:
Frequency and Frame rate settings may affect the Low Velocity Reject.
-
Adjust
Power
to obtain an acceptable image using the lowest setting possible. This is
particularly important in CW mode, as the energy duty cycle is 100% (constant).
NOTE:
The Doppler Power setting affects only Doppler operating modes.
-
Adjust the following settings to further optimize the display of the image.
-
Use the
Horizontal sweep
to optimize the sweep speed.
-
To view signal detail, adjust
Scale
to enlarge the vertical spectral Doppler trace.
-
Use
Invert
to reverse the vertical component of the spectral Doppler area of the display.
-
Use
Angle correction
to steer the ultrasound beam to the blood flow to be measured.
Summary of Contents for 5935000E10
Page 1: ...LOGIQ E10 Basic Service Manual 5750007 1EN Rev 1 Version R1 2018 GENERAL ELECTRIC COMPANY ...
Page 2: ......
Page 4: ...DIRECTION 5750007 1EN REVISION REV 1 LOGIQ E10 BASIC SERVICE MANUAL ii ...
Page 5: ...DIRECTION 5750007 1EN REV 1 LOGIQ E10 BASIC SERVICE MANUAL iii ...
Page 6: ...DIRECTION 5750007 1EN REVISION REV 1 LOGIQ E10 BASIC SERVICE MANUAL iv ...
Page 7: ...DIRECTION 5750007 1EN REV 1 LOGIQ E10 BASIC SERVICE MANUAL v ...
Page 8: ...DIRECTION 5750007 1EN REVISION REV 1 LOGIQ E10 BASIC SERVICE MANUAL vi ...
Page 9: ...DIRECTION 5750007 1EN REV 1 LOGIQ E10 BASIC SERVICE MANUAL vii ...
Page 10: ...DIRECTION 5750007 1EN REVISION REV 1 LOGIQ E10 BASIC SERVICE MANUAL viii ...
Page 11: ...DIRECTION 5750007 1EN REV 1 LOGIQ E10 BASIC SERVICE MANUAL ix ...
Page 12: ...DIRECTION 5750007 1EN REVISION REV 1 LOGIQ E10 BASIC SERVICE MANUAL x ...
Page 13: ...DIRECTION 5750007 1EN REV 1 LOGIQ E10 BASIC SERVICE MANUAL xi ...
Page 14: ...DIRECTION 5750007 1EN REVISION REV 1 LOGIQ E10 BASIC SERVICE MANUAL xii ...
Page 34: ...DIRECTION 5750007 1EN REV 1 LOGIQ E10 BASIC SERVICE MANUAL xxxii Hardware Kit 9 49 ...
Page 776: ...DIRECTION 5750007 1EN REV 1 LOGIQ E10 BASIC SERVICE MANUAL Index 4 ...
Page 777: ......