General Ultrasound system requirements
LOGIQ C9 Series – Basic Service Manual
2-11
5482298-100 English Rev. 5
Probes environmental requirements
Operation, storage and transport temperatures for probes
Probes should be operated, stored, or transported within the
parameters outlined below.
NOTE:
Refer to Table 2-1 on page 2-3 to determine the needed
settlement time.
Time and manpower requirements
Site preparation takes time. Begin site preparation checks as
soon as possible, if possible, six weeks before delivery, to allow
enough time to make any changes.
CAUTION
Ensure that the probe face temperature does not exceed the
normal operation temperature range.
Table 2-6: Probe Environmental Requirements
Operational
Storage
Transport
Temperature
10
°
- 40
°
C
50
°
- 104
°
F
0
°
- 55
°
C
32
°
- 131
°
F
-40
°
- 55
°
C
-40
°
- 131
°
F
Humidity
5 - 85%
non-condensing
5 - 85%
non-condensing
5 - 85%
non-condensing
Pressure
700 - 1060hPa
700 - 1060hPa
700 - 1060hPa
CAUTION
Check the room temperature before you use the probe.
CAUTION
Ensure that the probe face temperature does not exceed the
normal operation temperature range.
Summary of Contents for 5489329
Page 5: ...LOGIQ C9 Series Basic Service Manual i 3 5482298 100 English Rev 5 ...
Page 6: ...i 4 LOGIQ C9 Series Basic Service Manual 5482298 100 English Rev 5 ...
Page 7: ...LOGIQ C9 Series Basic Service Manual i 5 5482298 100 English Rev 5 ...
Page 8: ...i 6 LOGIQ C9 Series Basic Service Manual 5482298 100 English Rev 5 ...
Page 9: ...LOGIQ C9 Series Basic Service Manual i 7 5482298 100 English Rev 5 ...
Page 10: ...i 8 LOGIQ C9 Series Basic Service Manual 5482298 100 English Rev 5 ...
Page 11: ...LOGIQ C9 Series Basic Service Manual i 9 5482298 100 English Rev 5 ...
Page 12: ...i 10 LOGIQ C9 Series Basic Service Manual 5482298 100 English Rev 5 ...
Page 13: ...LOGIQ C9 Series Basic Service Manual i 11 5482298 100 English Rev 5 ...
Page 18: ...i 16 LOGIQ C9 Series Basic Service Manual 5482298 100 English Rev 5 ...
Page 52: ...Introduction 1 28 LOGIQ C9 Series Basic Service Manual 5482298 100 English Rev 5 ...
Page 76: ...Site Preparations 2 24 LOGIQ C9 Series Basic Service Manual 5482298 100 English Rev 5 ...
Page 118: ...System Setup 3 42 LOGIQ C9 Series Basic Service Manual 5482298 100 English Rev 5 ...
Page 206: ...Renewal Parts 9 8 LOGIQ C9 Series Basic Service Manual 5482298 100 English Rev 5 ...
Page 242: ...Care and Maintenance 10 36 LOGIQ C9 Series Basic Service Manual 5482298 100 English Rev 5 ...
Page 247: ...GE ...