When there's too much leakage current …
Venue 50 – Basic Service Manual
10-33
5447566-100 English Rev. 7
Probe Fails
Change another probe to confirm if the fail is caused by console.
NOTE:
Each probe will have some amount of leakage, dependent on its
design. Small variations in probe leakage currents are normal
from probe to probe. Other variations will result from differences
in line voltage and test lead placement. The maximum allowable
leakage current for body surface contact probe differs from
inter-cavity probe. Be sure to enter the correct probe type in the
appropriate space on the check list.
If excessive leakage current is slot dependent, inspect the
system connector for bent pins, poor connections, and ground
continuity.
If the problem remains with the probe, replace the probe.
Peripheral Fails
Inspect wiring for bad crimps, poor connections, or damage.
Still Fails
If all else fails, begin isolation by removing the probes, external
peripherals, then the on board ones, one at a time while
monitoring the leakage current measurement.
New Unit
If the leakage current measurement tests fail on a new
Ultrasound system and if situation can not be corrected, submit
a Safety Failure Report to document the Ultrasound system
problem. Remove Ultrasound system from operation.
Summary of Contents for 5448623
Page 5: ...Venue 50 Basic Service Manual i 3 5447566 100 English Rev 7 ...
Page 6: ...i 4 Venue 50 Basic Service Manual 5447566 100 English Rev 7 ...
Page 7: ...Venue 50 Basic Service Manual i 5 5447566 100 English Rev 7 ...
Page 8: ...i 6 Venue 50 Basic Service Manual 5447566 100 English Rev 7 ...
Page 9: ...Venue 50 Basic Service Manual i 7 5447566 100 English Rev 7 ...
Page 10: ...i 8 Venue 50 Basic Service Manual 5447566 100 English Rev 7 ...
Page 11: ...Venue 50 Basic Service Manual i 9 5447566 100 English Rev 7 ...
Page 12: ...i 10 Venue 50 Basic Service Manual 5447566 100 English Rev 7 ...
Page 22: ...i 20 Venue 50 Basic Service Manual 5447566 100 English Rev 7 ...
Page 52: ...Introduction 1 30 Venue 50 Basic Service Manual 5447566 100 English Rev 7 ...
Page 72: ...Site Preparations 2 20 Venue 50 Basic Service Manual 5447566 100 English Rev 7 ...
Page 160: ...Components and Functions Theory 5 4 Venue 50 Basic Service Manual 5447566 100 English Rev 7 ...
Page 204: ...Troubleshooting 7 38 Venue 50 Basic Service Manual 5447566 100 English Rev 7 ...
Page 228: ...Replacement Procedures 8 24 Venue 50 Basic Service Manual 5447566 100 English Rev 7 ...
Page 284: ...Care and Maintenance 10 36 Venue 50 Basic Service Manual 5447566 100 English Rev 7 ...
Page 288: ...Index 4 Venue 50 Basic Service Manual 5447566 100 English Rev 7 ...
Page 289: ......