Preventive maintenance:
Battery maintenance
5-14
Dash 3000/4000/5000
2000966-542D
5. Select
Wake Up Lithium Ion Battery in SLOT A
. Then, the
patient monitor applies a “wake-up” shot of current for up to 210
seconds.
6. Watch for the message,
Battery awake, run main code to charge
.
The battery should now be “awake.”
If the
“Batter
y awake, run main code to charge
” message
appears but the battery is NOT awake, the battery may be in a
deep discharge state. You should repeat step 5 at least four more
times to give the battery additional “wake-up” shots of current.
7. Quickly reboot (or power cycle) the patient monitor to begin charging
the batteries. The patient monitor’s illuminated yellow-colored
CHARGING STATUS
LED indicates that the battery is being
charged.
NOTE
The patient monitor will not charge the battery while it is
running the Boot Loader program.
How to replace the batteries
WARNING
EXPLOSION OR FIRE — Using non-recommended
batteries could result in injury/burns to the patient or
users.
Only use batteries recommended or manufactured by GE.
The warranty can be voided if non-recommended
batteries are used.
1. Open the battery doors on the left side of the patient monitor, along
the bottom.
2. Remove the faulty batteries.
Summary of Contents for 001C 051D 003A
Page 10: ...viii Dash 3000 4000 5000 2000966 542D ...
Page 11: ...2000966 542D Dash 3000 4000 5000 1 1 1 Introduction ...
Page 20: ...Introduction Service information 1 10 Dash 3000 4000 5000 2000966 542D ...
Page 21: ...2000966 542D Dash 3000 4000 5000 2 1 2 Equipment overview ...
Page 64: ...Equipment overview Theory of operation 2 44 Dash 3000 4000 5000 2000966 542D ...
Page 65: ...2000966 542D Dash 3000 4000 5000 3 1 3 Installation ...
Page 73: ...2000966 542D Dash 3000 4000 5000 4 1 4 Configuration ...
Page 99: ...2000966 542D Dash 3000 4000 5000 5 1 5 Preventive maintenance ...
Page 117: ...2000966 542D Dash 3000 4000 5000 6 1 6 Troubleshooting ...
Page 146: ...Troubleshooting Wireless LAN 6 30 Dash 3000 4000 5000 2000966 542D ...
Page 147: ...2000966 542D Dash 3000 4000 5000 7 1 7 Field replaceable units ...
Page 201: ...2000966 542D Dash 3000 4000 5000 8 1 8 Functional and electrical safety checks ...
Page 257: ...2000966 542D Dash 3000 4000 5000 A 1 A Electromagnetic compatibility EMC ...
Page 267: ...2000966 542D Dash 3000 4000 5000 B 1 B Network troubleshooting ...
Page 279: ...2000966 542D Dash 3000 4000 5000 D 1 D Checklist ...
Page 281: ......