Display
Action
Step
Select menu Check Lamp Run Time.
2
Flip time
The flip time is the time the monochromator stepper motor has been in operation.
Display
Action
Step
Select menu Check, press OK.
1
Select menu Check Flip Time.
2
If the flip time is more than 2000 hours, contact GE for maintenance.
6.6
Cleaning the flow cell and optical connectors
A clean flow cell and optical connectors are essential for ensuring the correct operation
of the UV-monitor.
WARNING
Hazardous chemicals or biological agents. When using hazardous
chemicals or biological agents, make sure that the entire system
has been flushed thoroughly with bacteriostatic solution (e.g., NaOH)
and distilled water before service and maintenance.
CAUTION
Hazardous chemicals or biological agents in UV flow cell. Make
sure that the entire flow cell has been flushed thoroughly with
bacteriostatic solution (e.g., NaOH) and distilled water, before service
and maintenance.
Monitor UV-900 Operating Instructions 28962214 AD
59
6 Maintenance
6.5 Checking the instrument