Using an SD Card
Proceed to step
.
7.
Do one of the following:
• To cancel the update to the boot code, press any key other than
Enter
.
The following messages are displayed:
Boot code not updated.
Can update later from service setup.
• To update the boot code, press
Enter
.
The following messages are displayed:
Programming Primary Boot
Programming Over
System will shut down
Press any key to continue...
Press any key to shut down the device.
8.
After the device shuts down, press
Power
to restart the device.
9.
On the
Resting ECG
screen, verify the software version has been updated.
104
MAC
®
5500/MAC
®
5500 HD
2046275-016B