Recording a Resting ECG
In either case, users may then do either of the following:
• Select
Continue
to continue (print the ECG).
• Select
Cancel
to cancel.
Resting ECGs
A resting ECG is the default mode of the ECG cart system, although you may change
this in the system configuration. This section describes how to record a resting ECG
and the available options.
Recording a Resting ECG
The following steps describe how to conduct a resting ECG.
NOTE:
To take a stat ECG, go directly to step
1.
Prepare the patient as described in
“Preparing the Patient” on page 51
2.
Select
Patient Data
and enter the patient data as described in
Information” on page 57
3.
Adjust the
Speed
,
Gain
and
Low pass filter
until the waveforms are configured
as desired.
4.
If the patient has a pacemaker, turn on
Pace Enhance
.
For more information, see
“ECG Options” on page 73
5.
Select
More
>
Printer Leads
to scroll through the leads or change the lead
format.
6.
When the waveforms are configured, press
ECG
to begin the acquisition.
A progress bar indicates the percentage of the data acquired. When the
acquisition is complete, one of the following occurs, depending on the setting of
the
Preview Before Analysis
option on the
Resting ECG Setup
window.
• If the
Preview Before Analysis
option is enabled, a preview of the 10–second
ECG is displayed. Continue with step
• If the
Preview Before Analysis
option is not enabled, the ECG data is analyzed
and printed after it is acquired. Proceed to step
7.
While reviewing the preview, do one of the following:
• Discard the reading and press
Cancel
. Begin again from step
.
• Wait for the menu options to change and then continue with step
8.
Use the options to change patients, to print a copy of the ECG, or to save,
transmit, or reanalyze the data.
For more information on each option, see
“Post-Acquisition Options” on page 75
.
72
MAC™ 2000
2053535-002K
Summary of Contents for MAC 2000
Page 30: ...Introduction 30 MAC 2000 2053535 002K ...
Page 44: ...Product Overview 44 MAC 2000 2053535 002K ...
Page 56: ...Preparing the Patient 56 MAC 2000 2053535 002K ...
Page 62: ...Entering Patient Information 62 MAC 2000 2053535 002K ...
Page 88: ...Arrhythmia Mode Recording 88 MAC 2000 2053535 002K ...
Page 90: ...RR Analysis 90 MAC 2000 2053535 002K ...
Page 94: ...RR Analysis 94 MAC 2000 2053535 002K ...
Page 104: ...Stress Testing 104 MAC 2000 2053535 002K ...
Page 182: ...Maintenance 182 MAC 2000 2053535 002K ...
Page 212: ...Technical Specifications 212 MAC 2000 2053535 002K ...
Page 224: ...224 MAC 2000 2053535 002K ...
Page 225: ......