Troubleshooting
Report Errors
This section addresses the following report error:
“ACI-TIPI Statement is not Included
on Report” on page 190
ACI-TIPI Statement is not Included on Report
If the ACI-TIPI statement is not displayed when expected, do the following:
• Verify the
ACI-TIPI
option is activated.
For information on activating the
ACI-TIPI
option, refer to
“Options Setup” on page
.
• Verify
ACI-TIPI
is enabled on the ECG.
For information, refer to
“Resting ECG Setup” on page 118
.
• Verify the information
ACI-TIPI
requires was entered.
The ACI-TIPI statement prints only if the patient’s gender, date of birth, and chest
pain indication are included in the patient information.
• Verify the patient is 16 years old or older.
The ACI-TIPI statement does not print for pediatric patients.
• Verify the original ECG was acquired in an electrocardiograph with the
ACI-TIPI
option.
If you attempt to print an ECG that was imported from an external device, the
cart does not generate an ACI-TIPI statement; it prints only if the statement was
saved as part of the ECG.
System Errors
The following table identifies some potential errors that may occur while you are
operating the system, the possible causes, and a recommended course of action
to resolve the error.
If performing the recommended actions does not resolve the problem, contact
authorized GE Healthcare service personnel.
Problem
Cause
Solution
This icon is displayed and the
battery LED is flashing.
System is operating from
the battery and the battery
charge is low.
Connect the system to an AC
outlet to charge the battery.
This icon is displayed and the
battery LED is not lit.
System is operating from
AC power and battery is not
installed.
Install a battery.
190
MAC™ 2000
2053535-002K
Summary of Contents for MAC 2000
Page 30: ...Introduction 30 MAC 2000 2053535 002K ...
Page 44: ...Product Overview 44 MAC 2000 2053535 002K ...
Page 56: ...Preparing the Patient 56 MAC 2000 2053535 002K ...
Page 62: ...Entering Patient Information 62 MAC 2000 2053535 002K ...
Page 88: ...Arrhythmia Mode Recording 88 MAC 2000 2053535 002K ...
Page 90: ...RR Analysis 90 MAC 2000 2053535 002K ...
Page 94: ...RR Analysis 94 MAC 2000 2053535 002K ...
Page 104: ...Stress Testing 104 MAC 2000 2053535 002K ...
Page 182: ...Maintenance 182 MAC 2000 2053535 002K ...
Page 212: ...Technical Specifications 212 MAC 2000 2053535 002K ...
Page 224: ...224 MAC 2000 2053535 002K ...
Page 225: ......