System Configuration
Stress ECG Setup Fields–Page 2 (cont'd.)
Field
Comments
J+x Point Formula
Determines the method that calculates the post J-Point.
Options are:
• 0 ms
• 10 ms
• 20 ms
• 40 ms
• 80 ms
• Rautaharju (default value)
• RR/16
The numeric values (0 ms—80 ms) add the selected number
of milliseconds to the J-point
Calculation (E, J point)
Determines when the select J+x point formula is used. Valid
options are:
• Single
The E and J points are calculated once in the beginning and
remain unchanged during the stress test.
• Continuous
The E and J points are continuously updated during the
PRETEST, EXERCISE, and RECOVERY phases of the stress test.
136
MAC™ 2000
2053535-002K