Maintenance and checkout
131
2062973-002
10.2 Electrical safety checks
Electrical safety tests provide a method of determining if potential electrical health hazards to
the patient or operator of the device exist.
Perform the following electrical safety tests described in detail in chapter
:
−
−
−
−
8.2.4. Ground (earth) integrity
−
8.2.5. Earth leakage current test
−
8.2.6. Enclosure leakage current (touch current) test
Record the values of the tests on the Appendix
.
10.3 Functional check
10.3.1 Start-up
Follow the procedure in section
.
10.3.2 Display
Follow the procedure in section
10.3.3
PSM / PDM identification
1.
Configure the ECG1 waveform field and the NIBP parameter window to the patient
monitor screen with adequate priority. Connect a PSM or PDM module to the patient
monitor.
2.
Verify that the ECG waveform field, the NIBP parameter window and the related
information appear on the patient monitor screen.
10.3.4
E-module identification
1.
Log in to Webmin.
2.
Select
Information > Device Information
.
3.
Verify that the information about the connected module appears in the table named as
“Acquisition Information – E-Modules”
.
NOTE: For some parameter modules, the table does not show the actual module information,
but the information of the individual subassemblies inside the module.
NOTE: You may need to refresh the Webmin screen if you have connected the module to the
patient monitor after entering the
Device Information
Webmin screen.
10.3.5 Keypad and remote
Follow the procedure in section
10.3.6 Mouse
Follow the procedure in section
Summary of Contents for CARESCAPE Monitor B650
Page 12: ...CARESCAPE Monitor B650 4 2062973 002 For your notes ...
Page 16: ...CARESCAPE Monitor B650 8 2062973 002 For your notes ...
Page 46: ...CARESCAPE Monitor B650 38 2062973 002 For your notes ...
Page 58: ...CARESCAPE Monitor B650 50 2062973 002 For your notes ...
Page 64: ...CARESCAPE Monitor B650 56 2062973 002 For your notes ...
Page 136: ...CARESCAPE Monitor B650 128 2062973 002 For your notes ...
Page 177: ...Troubleshooting 169 2062973 002 ...
Page 182: ...CARESCAPE Monitor B650 174 2062973 002 For your notes ...
Page 218: ...CARESCAPE Monitor B650 210 2062973 002 For your notes ...
Page 224: ...CARESCAPE Monitor B650 216 2062973 002 For your notes ...
Page 232: ...CARESCAPE Monitor B650 B 4 4 2062973 002 For your notes ...
Page 239: ......